Ch.4 Checking Image Quality
< 145 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
2
Print out one internal densitometer calibration
sheet.
1)Invoke the menu mode by touching the menu key.
2)log-in to the service maintenance mode using
service privilege.
3)Touch the [CORRECTION] button under the
[SERVICE 2] item.
- The densitometer calibration/previous density
value selection screen will be displayed.
4)Touch the [DENSITOMETER CORRECT] button.
- The tray selection/print screen will be displayed.
5)Select the supply tray used for calibration in step-1
and touch the [PRINT] button.
- A densitometer calibration sheet will be printed.
- Proceed to step-3 without changing the screen.
(00)
(01)
(02)
(03)
(04)
(05)
(06)
(07)
(08)
(09)
(10)
(11)
(12)
3
Measure the densities shown in frames (01) -
(12) on the printed densitometer calibration
sheet.
- Record the densities measured either next to the
frames on the sheet or on a separate sheet of
paper.