Ch.6 Web Maintenance Mode
< 187 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Item
Default
Descriptions
STAMP SETUP
CH01 : OFF
This setting determines whether or not film is printed with a date
< TIME/DATE >
CH02 : ON
stamp.
CH03 : OFF
"ON"
: Date is printed.
CH04~16 : ON
"OFF"
: Date is not printed.
STAMP SETUP
CH1 : OFF
This setting determines whether or not film is printed with a
< SCU ID/PAGE NO.>
CH2 : ON
diagnostic device ID and page number stamp.
CH03 : OFF
Note, however, that, even when set to ON, these stamps are not
CH04~16 : ON
printed on small size film (10"X12", 8"X8").
"ON"
: The diagnostic device ID and page number are printed.
"OFF"
: The diagnostic device ID and page number are not printed.
STAMP SETUP
CH1 : OFF
This setting determines whether or not film is printed with a print
< COPY NO. >
CH2 : ON
count stamp.
CH03 : OFF
"ON"
: Print count is printed.
CH04~16 : ON
"OFF"
: Print count is not printed.
STAMP SETUP
CH1 : OFF
This setting determines whether or not film is printed with
< STAMP MESSAGE >
CH2 : ON
a message/print count stamp.
CH03 : OFF
If set to ON, the message may be set from the "stamp message"
CH04~16 : ON
setting.
"ON"
: Message/print count is printed.
"OFF"
: Message/print count is not printed.
STAMP SETUP
CH1 : OFF
This setting determines whether or not film is printed with a patient
< PATIENT ID / NAME>
CH2 : ON
ID/name stamp.
CH03 : OFF
"ON"
: Patient ID is printed.
CH04~16 : ON
"OFF"
: Patient ID is not printed.
STAMP SETUP
CH1 : OFF
This setting determines the print format of the patient ID and name.
< PATIENT DATA >
CH2 : ON
Valid only when the [Stamp Setup] <PATIENT ID> is set to ON.
CH03 : OFF
"ASCII"
: Alphabetical input.
CH04~16 : ON
"2BYTE"
: 2 byte character input.
ENLARGED CHARACTER
CH01~16 : OFF
Sets the stamp (patient name) print size.
"ON"
: The patient information is printed in enlarged format.
"OFF"
: The patient information is printed in normal size.
ex)
STAMP MESSAGE
CH01~16 : (Blank)
Regardless of whether full or half size characters are used, a
maximum of 32 characters may be input as the message printed on
the stamp.
In cases where characters other than alphanumericals are set in the
Web maintenance mode, the message content may be referenced,
but not changed. Changes may be effected from the Web
maintenance mode.
Valid only when the [Stamp Setup] <STAMP MESSAGE> is set to ON.