Ch.1 Pre-installation Information
< 18 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Connected to CS3 (Venus PCM Mode)
CS-3
DRYPRO 793
14x17 BLUE Dmax3.0
11x14 DR BLUE Dmax4.0
8x10 DR BLUE Dmax4.0
Tray-1 : 14x17 B
Tray-2 : 11x14 M
Tray-3 : 8x10 M
SCU Setting
Size Type Check
DRYPRO 793
ON
1.2.3
25
µ
Output
For a Regius-PCM system, the DRYPRO 793 super-fine mode (25
µ
m) is required.
For connections to Regius contact mammography/normal photography and other modalities, the
normal mode (43.7
µ
m) is used.
DRYPRO 793 requires the conditions shown below for 25
µ
m output.
- 1GB expanded print memory.
- Loaded with mammography-dedicated film (film type <M>).
An example of output in the 25
µ
m mode is shown below.
For 25um output, a command for output on the "DR BLUE FILM" film type must be
sent from CS-3.