Ch.1 Pre-installation Information
< 19 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Connection to Devices of other Manufacturers Using Printlink
The "BLUE FILM" output command from devices of other manufacturers must be converted to
"DR BLUE FILM" (Konica Minolta's unique standard) and sent to DRYPRO 793. For details of he
method of conversion and other relevant information, refer to the Printlink Manual.
Printlink
DRYPRO 793
14x17 BLUE Dmax3.0
11x14 DR BLUE Dmax4.0
8x10 DR BLUE Dmax4.0
14x17 BLUE Dmax3.0
11x14 BLUE Dmax4.0
8x10 BLUE Dmax4.0
Device of Other
Manufacturer
Tray-1 : 14x17 B
Tray-2 : 11x14 M
Tray-3 : 8x10 M
SCU Setting
Size Type Check
DRYPRO 793
ON
1.2.4
Mammo-film Output
In mammo-mode of other manufacturers, DRYPRO 793 high-density output (Dmax 4.0) is
required. For Regius normal photography and other modalities, Dmax 3.0 is used.
DRYPRO 793 requires the conditions shown below for Dmax 4.0 output.
- Loaded with mammography-dedicated film (film type <M>).
An example of mammo-film output is shown below.
Connection with Devices of other Manufacturers without Printlink
To send a direct output command from the device that cannot change the Dmax, the following
settings must be made in DRYPRO 793.
[PRINT SETUP]
[PRINT CONDITION]
[Maximum density]
: "400"
[Maximum/minimum density value used] : "DICOM"
[Parameter priority setting]
: DENSITY MAX set to ON.
DRYPRO 793
14x17 BLUE Dmax3.0
11x14 BLUE Dmax4.0
8x10 BLUE Dmax4.0
Device of Other
Manufacturer
Tray-1 : 14x17 B
Tray-2 : 11x14 M
Tray-3 : 8x10 M
SCU Setting
Size Type Check
DRYPRO 793
OFF
p.110 PRINT
CONDITION
SETUP