Ch.1 Pre-installation Information
< 22 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
1.3.2
Checking the Network
The following points should be checked with the facility network administrator in advance of
installation.
DRYPRO 793 (SCP) Related Settings
- IP addresses and sub-net masks (when DHCP server is not used) of DRYPRO
793, maintenance PC and gateways.
- DRYPRO 793 host name.
- DRYPRO 793 AE title and port number (print service).
Diagnostic Device Related Settings
- Number of diagnostic devices connected to DRYPRO 793.
- IP address of each diagnostic device.
- AE title of each diagnostic device.
- Port number of each diagnostic device (N-EVENT REPORT service).
- Print conditions used by each diagnostic device (format, film size).
Maintenance PC
- Model with 10BaseT/100BaseTX/1000BaseT network ports.
- PC with TCP/IP protocol installed.
- PC with Windows2000/XP OS.
- PC with Internet Explorer (version 5.5 or higher).
- PC with environment capable of executing Java script with the Web browser.
Equipment for Network Connection
- Hub (Used for direct connection to DRYPRO 793 and maintenance PC).
Hub with auto-negotiation capabilities as 100Base-TX hub recommended.
- Ethernet Cables
(For DRYPRO 793 -> Hub connection and maintenance PC -> Hub connection)
CAT5E or CAT6 should be used for 1000Base-T.
Tools
- Box Cutter
- Phillips screw driver
- Long phillips screw driver (90mm or longer) for disengaging the front cover lock.
- Flat head screw driver
- Radio pliers
- 13mm box wrench (with a long section below the neck). Alternatively, socket
wrench (with extension cover)
- Portable densitometer: X-Rite 331 (used to check image)
- Magnifying glass (used to check image)
1.3.3
Equipment and Tools Required for Installation
Before proceeding with installation, check to ensure that the following equipment and tools are
available.
p.93 SCP SETUP
p.96 SCU SETUP