3 .2 Image Erase Operation
31
Chapter 3
6
Wait for the image erase process to
complete .
CAUTION
•
Do not touch the device, the inserted cassette, or the
platform/table the device is set on while images are
being erased .
•
When the image erase process has completed,
the Ready Lamp (green) will start blinking to in-
dicate that the cassette can be removed .
7
Confirm that the Ready Lamp is blinking,
then remove the cassette .
•
When removing the cassette, the device initial-
ization process will begin, and the Busy Lamp
will be lit .
•
Once the initialization process has ended, the
Busy Lamp will turn off, and the Ready Lamp
will be lit .
IMPORTANT
•
Do not insert the cassette during the initialization pro-
cess (while the Busy Lamp is lit) . Doing so will cause
initialization to end abnormally, and an error will occur .
HINT
When an image remains after erasing once
•
When the image erase process was performed once
for an image exposed at high dosage, etc., but was
not completely erased, the second image erase pro-
cess will be performed .
There are two ways to execute the second erase pro
-
cess: “Auto Run” and “Display message then se lect
run” . The setting can be changed by the user tool .
Refer to the Image processing controller Operation
Manual .
When responding to a message to perform the
second erase process
•
If you need to perform the second erase process, the
Error Lamp (yellow) will light up, and an image re-
erase request message will be displayed in the Image
processing controller .
•
Clicking the [Continue] button at the message will start
the image erase process .
• Clicking the [Eject] button will eject the cassette, and
the second erase process will not be performed . When
doing this, make sure to perform the image erase pro-
cess at a later time .
•
Even when selecting the message display option, the
second erase process will be performed automatical ly
if Image processing controller has not started up or is
not connected to the device .
Summary of Contents for REGIUS SIGMA2
Page 1: ...EN 05 0197 Operation Manual DIRECT DIGITIZER REGIUS SIGMA2...
Page 2: ......
Page 5: ...5 Introduction...
Page 10: ...10...
Page 21: ...21 Chapter 2 Product Outline An explanation of the device s features and functions...
Page 27: ...27 Chapter 3 Reader Operation An explanation on how to operate this device...
Page 35: ...35 Chapter 4 Cassette Plate Operation An explanation on cassette and plate handling...
Page 45: ...45 Chapter 5 Troubleshooting An explanation on how to deal with errors when they occur...
Page 62: ...62...
Page 70: ...70...
Page 71: ...71 Appendix A list of device specifications etc...
Page 74: ...74...
Page 75: ......
Page 76: ...A5ERBA01EN05 2015 08 31 JD...