SKR 4000 DIRECT DIGITIZER User Manual
50
..................................................................................................................................................KONICA
MINOLTA, INC.
7.
X ray generator starts releasing x ray
8.
X ray generator finishes x ray shooting.
9.
Workstation receives “Acquire” request, send command “Data Acquisition” to panel.
10. Panel receives “Data Acquisition” from workstation, start data acquisition operation.
11. Panel completes image acquisition and begins to send data to workstation.
12. Workstation receives all image data from panel which are after calibration if Hardware calibration
is on.
4.1.2.
Inner2 Mode
4.1.2.1. Block Diagram
Workstation(Inner2)
HVG
Tube
Workstation is a host PC device installed with iDetector and SDK. Chapter 3 has described how to
establish connection between panels and workstation. In inner mode, workstation does not control x ray
generator. Users would decide when to shoot x ray.