background image

 

ViZion DR+ 1417CK Wireless Calibration Guide v1.0 

Ultra UAI v4.1.0.X

 


 

 

 

 

 

 

 

           

Customer Support 

1.800.366.5343 – 

HCITsupport@konicaminolta.com

 

VZ0148UG201611-1.0 

 

 

3.  Click 

General Settings

 on the left 

menu, set the following general 
configuration settings. 

Trigger

 = Inner Trigger 

Exp Window

 = 30000 

 

Figure 11: iDemo General Settings 

4.  Click 

Write RAM

 to save changes. 

5.  Click 

HW Correction

 settings on the left 

menu and set the following setting. 
 

Offset Mode

 = No 

 

Figure 12: HW Correction Settings 

6.  Click 

Write RAM

 to save changes. Gain 

and Defect Map settings changes to “Do 
in SW” (Figure 12). 

7.  Click 

Exit

 to exit the Config settings 

window (Figure 12). 

3  Calibrating ViZion DR+ 

3.1  Environment Setup 

1.  Align the panel with the X-ray tube. 

2.  Set tube head Source to Image-receptor 

Distance (SID) to 

60”

.  

3.  Open the collimator to allow full 

coverage of the surface of the panel. 
The collimator light should extend an 
inch beyond the edges of the panel. 

4.  Remove any anti-static grid or grid-cap. 

Make sure panel is out of bucky and no 
objects are between the panel and tube 
head. 

5.  Configure 

70 kV 5.4 mAs

 on the X-ray 

control console. 

3.2  Performing Panel Gain 

Calibration 

If using an integrated generator, it must be 
configured to make exposures in Direct Mode to 
run calibration. Use the manufacturer software 
parallel to iDemo to control the generator 
settings, or adjust kV and mAs from the 
generator console. 

1.  In the Correction section, configure the 

following setting. 

 

Offset

 = POST 

 

Figure 13: Correction Settings 

Summary of Contents for ViZion DR+ 1417CK

Page 1: ...ViZion DR 1417CK Calibration Guide...

Page 2: ...ion DR 1417CK 2 1 Preparing for Calibration 2 1 1 Recommended Panel Configurations 2 1 2 Powering on the Panel 2 1 3 Stopping the Maven Process 3 2 iDemo 3 2 1 Connecting to iDemo 3 2 2 Configuring iD...

Page 3: ...anel KV300XXXXXXXX 1 Preparing for Calibration 1 1 Recommended Panel Configurations Prior to starting panel calibration it is recommended that basic device configuration is setup properly in the ViZio...

Page 4: ...easing Figure 4 Power Button 1 3 Stopping the Maven Process 1 Exit the ViZion DR acquisition software 2 Stop the Maven process if it is running a Verify if Maven is running An iRayMk3 icon is displaye...

Page 5: ...ow Do not close Figure 7 FTP Utility 4 Launch the iDemo calibration utility a Navigate to c IDemo_3 0 0 14 b Double click the file IDemo_interface exe 5 Click Start Figure 8 iDemo Connect Detector 6 I...

Page 6: ...ng ViZion DR 3 1 Environment Setup 1 Align the panel with the X ray tube 2 Set tube head Source to Image receptor Distance SID to 60 3 Open the collimator to allow full coverage of the surface of the...

Page 7: ...s at the bottom of the screen While the image is transferring the status changes from Image Getting showing the panel has triggered an image to Exposure prohibit post offset to Exposure enable post of...

Page 8: ...add that value to the Selected Points list Figure 18 Figure 17 Center Average Value Figure 18 Selected Points List 10 Repeat Steps 5 6 until you have a total of 5 gain frame values in the Selected Po...

Page 9: ...ect tab The Current Frame defaults to 1 A total of 19 frames must be acquired to complete a full defect mapping Figure 22 Defect Calibration 4 Configure the dose as close to the requested kV as possib...

Page 10: ...f the expected gray value for that frame If the value is more than 100 below or more than 100 above the expected gray value adjust your mAs up or down NOTE If the Center value returned is too low or t...

Page 11: ...quent frames 5 19 when you click Start Creating after frame 4 12 After obtaining 19 19 frames click Create Defect Figure 27 Creating Defect File 13 Click OK when the Correction file is generated succe...

Page 12: ...tus to change to Download Hardware Correction File Success 6 Click Write RAM to save changes Figure 29 HW Correction 7 Click Exit to exit the Config Screen 8 On the Correction section a Select dropdow...

Reviews: