Examination screen
Messages
Causes and corrective actions
Darken the room
The room is deemed to be too bright to perform an
examination. Adjust the luminance of the room to be the
same as that when the instrument is started.
Can not recognize the patient's
eye. Please make sure that the
chin is placed correctly on the chin
rest.
Do you want to manually select the
patient's eye?
[Start] has been pressed when the patient
’
s chin does not
yet come into contact with the chin rest. Confirm that the
patient
’
s chin is surely placed on the chin rest.
Press [OK] to manually specify the eye to be examined.
Press [Cancel] to close the message box.
Failed to acquire the pupil position
Do you wish to proceed with
acquisition from the beginning?
You have failed in gaze monitor setting.
Pupil position acquisition failure can be caused by
eyelashes, eyelid ptosis, opacity of the middle translucency
body, and small pupil.
To retry the acquisition, confirm the eye fixation, and then
press [OK]. Not to retry it, press [Cancel].
If you press [Cancel], the gaze monitor function will be
deactivated for this examination.
The fovea is examined first.
The fixation target is the 4 points at
the lower center.
[Start] has been pressed on the examination screen at the
time of examination startup with the fovea examination set
to ON. The fixation target is automatically moved to 4 points
at the lower center. Press [OK] to start the fovea
examination.
Next, the threshold examination
starts
(Next, the isopter examination
starts)
The fixation target is the selected
one
Fovea examination has been completed with the fovea
examination set to ON. The fixation target is automatically
returned to the original location. Press [OK] to start the
threshold (isopter) examination.
Put back the lens frame.
This message appears when the lens holder is set for the
periphery examination (more than 30
°
) or upon completion
of center examination with the lens holder set in the
program consisting of both center (within 30
°
) and
periphery examinations. Thus, be sure to fold the lens
holder backward.
In the program consisting of both center and periphery
examinations, the stimulus is normally presented in random
order. However, when the lens holder is set, the periphery
examination is carried out upon completion of the center
examination. Therefore, this message appears after the
center examination is completed.
Please switch eyes
Attempt has been made to re
–
test an eye other than the
one examined previously. Check the eye to be examined:
left or right.
Select the point examined again
[OK] has been pressed without examined points selected
for re
–
testing. Select points to be re
–
tested.