GG-11008 Edition 1.2
- 19 -
Messages
Causes and corrective actions
Next, the screening examination
starts
(Next, the threshold examination
starts)
Isopter examination has been completed during the Isopter
+ Screening (threshold) examination program. Press [OK]
to start the screening (threshold) examination.
The examined eye is incompatible
with the fundus image
During perimetry coordinate setting on a fundus image for
the fundus perimetry, the positions of the fovea and opti
disc have been reversed in comparison to the actual
examined eye. Check whether the examined eye matches
the fundus image. In addition, check whether the examined
eye is left or right.
Maximum number of isopters
reached
Attempt has been made to start the Isopter
–
Manual
examination with 6 types of stimulus. Up to 5 types of
stimulus can be selected for an isopter examination. To
perform an examination with 6 or more types of stimulus,
complete one examination with some types of stimulus
once, and then, perform another examination with the other
types of stimulus.
Result screen
Messages
Causes and corrective actions
The combined data is not saved.
Continue operation?
Attempt has been made to change over the screen to
another without saving the result of data combination. To
save the combined data, press [Cancel] and [Save] in this
order. Not to save, press [OK] and proceed with the
operation.
The data has been properly copied Data read has been normally completed from Kowa
automatic perimeter
“
AP
–
6000,
”
“
AP
–
5000,
”
or
“
AP
–
3000
”
to
“
AP
–
7000.
”
The remained memory of the drive
is less than 5%
This message appears when the remaining free space of
the data save drive is less than 5 at the time of software
startup or examination result save.
Database/network errors
Messages
Causes and corrective actions
Network connection error
Network drive has been shut
down.
Could not connect to the network
If you proceed there will be no
link to the VK (any) database.
Cannot open the database
If the examination continues, the
results will be saved in the local
folder separately from previous
data.
The database or network is abnormal in some way. Select
[OK] to save the data locally. If the database or network is
restored when the instrument is started next time, the
locally saved data can be coupled to the conventional data.
Select [Cancel] to end KOWA AP
–
7000.