GG-11008 Edition 1.2
- 35 -
6-3.3.1.2 Origin position setting
Adjust the two motors (horizontal and vertical) and set the origin.
1) Each motor moves to the set origin. The screen shows the motor setup values at the center.
The stimulus lamp goes lit and the shutter opens. Left std. is selected.
2) Using the right and left arrow buttons on the right side of the screen, adjust the stimulus so
that it is positioned at the dome center LED while checking visually.
3) Pressing Position gap check moves the stimulus lower left and returns to the origin. Check
that the origin is positioned correctly.
4) Press Right std. button to present the stimulus at the origin with the right reference. Perform
steps 2) and 3) in the same way as for the left reference.
Figure 58 Origin position setting