GG-11008 Edition 1.2
- 46 -
(2) Stimulus maximum intensity
Standards
Test procedure
Tools
(Standard value) ±20%
See the table below for
standard values and
tolerances
Perform the test in a dark room or cover the dome
front side with a blackout curtain for the test.
Perform this test after setting the intensity (see
Section 3.3.1.4 "Stimulus brightness setting". The
measurement intensity [dB] and positions are listed
below. According to this table, set the intensity using
the stimulus intensity up/down arrow buttons, project
the stimulus to the related positions using the
stimulus projection position button, measure the
intensity with the intensity meter and check that the
results satisfy the standards.
Intensity
meter
Stimulus measurement intensity, measurement points, standard values and tolerances
Intensity [dB]
Measurement
positions
Standard values
[cd/m
2
]
MIN[cd/m
2
]
MAX[cd/m
2
]
0
A, B, C, D, E
3,183
2,546
3,820
Stimulus positions (polar displays)
Points
r[°]
θ
[°]
A
4
90
B
20
135
C
20
225
D
20
315
E
20
45
(3) Intensity changes
Standard
Test procedure
Tools
(Standard value) ±20%
See the table below for
standard values and
tolerances
Perform the test in a dark room or cover the dome
front side with a blackout curtain for the test.
Perform this test after setting the intensity (see
Section 3.3.1.4 " Stimulus brightness setting ". The
measurement intensities [dB] and positions are listed
below. According to this table, set the intensity using
the stimulus intensity up/down arrow buttons, project
the stimulus to the related positions using the stimulus
projection position button, measure the intensity with
the intensity meter and check that the results satisfy
the standards.
Intensity
meter