7 Connection
7.3 Connecting the pick up
2.
Use the diagonal cutter
4
to adapt the cable entry
3
to the cable diameter.
3.
Pull the cable
5
through the cable entry
3
.
6
2
2
7
1
4.
Hook the short arbour of the terminal tool
2
into the tension spring terminal
6
and press away
from the cable opening so that the cable opening opens.
5.
Insert the wire
7
into the cable opening and remove the terminal tool
2
.
6.
Repeat Steps 4 and 5 for all the wires.
7.
Replace the rear device cover
1
.
7.3 Connecting the pick up
Personnel qualification:
o
Electrician
Personal protective equipment:
o
Work clothing
Aids:
o
KRAL tool set
o
Diagonal cutter
o
Wiring diagram
ATTENTION
Device damage through incorrect connection
► Observe pin assignment and connection data of the electronic unit
► Before connecting the electronic unit to the power supply, ensure that all consumers (sensors) are
connected correctly. See the wiring plan.
Requirement:
ü
Pick up for flowmeter installed
1.
Remove the rear device cover.
2.
Use the diagonal cutter to adapt the cable entry to the cable diameter.
3.
Pull the cables of the pick ups individually through the cable entries.
4.
Connect the cables for pick ups of flowmeter A in accordance with the wiring diagram on the
electronic unit.
5.
Replace the rear device cover.
7.4 Connecting analog outputs and pulse inputs
Personnel qualification:
o
Electrician
Personal protective equipment:
o
Work clothing
Aids:
o
KRAL tool set
o
Diagonal cutter
o
Wiring diagram
Operating instructions
OIE 15en-GB Edition 2020-03
15