Installation and operating instructions BM 702 (07/00)
2.2
Mechanical installation
Hazardous-duty systems:
•
The BM 702 Ex is certified in conformity with
European Standard
for use in Zone 0, 1
and 2 hazardous locations (dependent on version).
•
Attention is drawn to the data and information given on the
nameplate of the
converter
, the
nameplate of the flange
and the specifications in the
approval
certificates
.
Safety:
•
Check
material compatibility
of antenna, extension, flange, gaskets, and PP or PTFE
(used in all versions) with the product! See also section 8 "Type code"!
Mounting on the tank nozzle
a) Devices with horn antenna:
The antenna should project out of the nozzle. If necessary, use an antenna extension.
Exception: in case of a symmetrical tank fitting.
Tank nozzle
Tank nozzle
Antenna
extension
Tank nozzle
b) Wave-Stick
Note the requirements imposed on nozzle diameter and nozzle length:
min. 50 mm
min. 200 mm
Tank nozzle
e.g. DN50
dia. 40...55 mm
Version for
still wells