3.6 Connecting the device
•
Select the adapter cable suitable for your device.
•
Plug in the USB connector
2
into the USB port of the primary master
1
.
•
Connect the M8 cable connector (male)
4
with the M8 cable connector (female)
3
and tighten
it by hand (max. 5 Nm).
•
Screw the VP cable connector
5
on the sensor connector
6
and tighten it by hand
(max. 5 Nm).
WARNING!
During installation of the device make sure that you use ESD (electrostatic discharge) protection
equipment.
CAUTION!
The resistor (load) for HART
®
communication is already integrated in the OPTIBRDIGE.
CAUTION!
Moisture on the sensor connector must be avoided! Moisture may cause a short-circuit and a
malfunction of the sensor!
If moisture has entered the connector dry it with air (e.g. hot air gun).
Figure 3-2: Connecting with adapter cable M8-VP2
1
Primary master (PC / notebook with PACTware
TM
FDT/DTM)
2
USB-connector
3
M8 cable connector (female)
4
M8 connector (male)
5
VP cable connector
6
Sensor connector