For initial set up, we strongly recommend that you also refer to the relevant manual!
Quick Start
1
Installation, assembly, start-up and maintenance may only be performed by appropriately
trained personnel. Check the nameplate for correct operating conditions.
For use in hazardous areas, special codes and regulations are applicable. Instruments must
not be connected to power supply before reading instructions described in the supplementary
manual.
This instrument complies with the requirements of the Pressure Equipment Directive. Please
refer to the nameplate for operating condition limits. Instruments must not be pressurised
before reading instructions described in the manual.
This instrument complies with the requirements of the Low Voltage Directive. Instruments must
not be connected to the power supply before reading the instructions described in the manual.
For devices used in SIL applications, additional safety notes apply. For detailed information
refer to the “Safety Manual”.
The responsibility as to the suitability, intended use and corrosion resistance of the used
materials against the measured
fl
uid of this device rests solely with the operator.
For complete documentation (manuals, supplementary manuals, data sheets and certi
fi
cates)
please refer to
www.krohne.com/Downloads
.
1 Installation
Quick Start
Hygienic connections
Flange connections
Handling the meter
Supporting the meter
Mounting positions
Self draining (hygienic)
Special conditions of use to be observed
• The OPTIMASS measuring system and MFC converter must be included in the equipoten-
tial bonding system of any hazardous area installation.
• Allow the electronics to de-energise before opening the electronics enclosure inside a
hazardous area. The time needed to de-energise is 35 minutes for temp. class T6 and 10
minutes for Temp. class T5. There is no waiting time for Temp. class T1
• Use certi
fi
ed cable glands on cable entries. Use a certi
fi
ed plug to seal unused cable
entries.
• Install connecting cables as
fi
xed wiring so that they are protected against damage.
Connecting cables must not exceed the parameters speci
fi
ed in the Supplementary
Instructions for Hazardous Areas.
•
For details on the relationship between: maximum permissible ambient temperature, maxi-
mum medium temperature, maximum surface temperature and the relevant temperature
class for each individual meter, please refer to the tables provided in the relevant meter
Handbook and/or the Supplementary Instructions for Hazardous Areas.
If the hygienic installation requires the meter to
be self draining, mount the meter at an angle.
DO NOT mount the meter horizontally!
04/2020 - 4008121701- QS-OPTIMASS 1400-R01- en
OPTIMASS 1400
Twin, straight tube mass
fl
owmeter
1