DEVICE DESCRIPTION
2
15
OPTISWIRL 4200
www.krohne.com
08/2016 - 4005002601- AD OPTISWIRL4200 Ex-II2GD R01
2.9 Electrical data
Signal circuits
Signal circuits
Signal circuits
Signal circuits
The vortex flowmeter signal circuits may only be connected to circuits with the following
maximum values. A U
m
of 253 V is required for the power supply units.
Flow sensor circuits
Flow sensor circuits
Flow sensor circuits
Flow sensor circuits
With the compact device, the intrinsically safe flow sensor circuits are designed as internal
circuits.
When it comes to the remote versions, the intrinsically safe flow sensor circuits are led through.
The maximum permissible values of the flow sensor circuits considered safe are listed below:
Pressure sensor circuit (connector X2.1...X2.5)
Pressure sensor circuit (connector X2.1...X2.5)
Pressure sensor circuit (connector X2.1...X2.5)
Pressure sensor circuit (connector X2.1...X2.5)
•
U
o
= 6.65 V
•
I
o
= 35 mA
•
P
o
= 58 mW
•
C
o
= 2.5
µ
F
•
L
o
= 0.5 mH
Piezo / Temperature sensor circuit (connector X2.7...X2.12)
Piezo / Temperature sensor circuit (connector X2.7...X2.12)
Piezo / Temperature sensor circuit (connector X2.7...X2.12)
Piezo / Temperature sensor circuit (connector X2.7...X2.12)
•
U
o
= 6.65 V
•
I
o
= 21 mA
•
P
o
= 35 mW
•
C
o
= 2.6
µ
F
•
L
o
= 0.5 mH
Device version
Circuit
Terminals
Nominal voltage
Nominal current
OPTISWIRL 4200 C Ex
VFC 200 F 020 Ex
Current output 4...20 mA
C1, C2
12...32 VDC
4...20 mA
Binary output
M1, M2, M3, M4
8...32 VDC
≤
100 mA
Current input
I1, I2
9...32 VDC
0...20 mA
OPTISWIRL 4200 C FF Ex
VFC 200 F FF 020 Ex
1
Bus connection
D, D-
9...32 VDC
20 mA
OPTISWIRL 4200 C PA Ex
VFC 200 F PA 020 Ex
2
1
Further information for operation of the FF transmitter are provided in separate supplementary instructions.
2
Further information for operation of the PA transmitter are provided in separate supplementary instructions.