OPERATION
5
23
OPTISWIRL 4200
www.krohne.com
08/2016 - 4005002601- AD OPTISWIRL4200 Ex-II2GD R01
5.1 Start-up
Start-up may only initiate when the vortex flowmeter:
•
is correctly installed in the system and connected.
•
has been checked for the proper state with regard to its installation and connection
requirements.
•
has been correctly locked to the electronics and terminal compartment.
The user of the system must have it checked before start-up in compliance with the national
regulations for checks before startup.
If the device needs to be configured due to the existence of an explosive atmosphere, this can be
done using the supplied bar magnets. There is no need to open the housing as it can be done
through the glass window of the electronics compartment or digitally via the signal output
(HART
®
interface).
5.2 Operation
Vortex flowmeters must be operated in such a way that they remain within the maximum and
minimum permissible temperatures and pressures and the electrical limit values.
Vortex flowmeters may only be operated if the equipment parts necessary for safety are effective
in the long run, and are not rendered inoperable during operation.
In case of operation with flammable products the measuring units must be included in the
periodic pressure tests of the system.
Opening the signal converter (protection type flameproof enclosure or dustproof housing or
terminal compartment in increased safety) in hazardous areas is only permitted in a de-
energised state.
5.3 Electrostatic charge
In order to avoid ignition hazards due to electrostatic charge, vortex flowmeters may
not be used in areas with:
•
processes that generate strong charges,
•
mechanical friction and cutting processes,
•
spraying of electrons (e.g. in the vicinity of electrostatic painting systems) or
•
pneumatically conveyed dust is exposed.
CAUTION!
Electrostatic charging of the housing surface by friction must be avoided. The devices must not
be dry cleaned.