INSTALLATION
3
29
OPTISWIRL 4200
www.krohne.com
02/2022 - 4003930810 - MA OPTISWIRL4200 R10 en
Figure 3-3: Not recommended installation
1
Installing the device in a downstream pipe
2
Installing the device in front of an outlet
3
Installing the device in an upper pipe bend due to risk of gas bubbles forming
CAUTION!
•
Installing the device in a downstream pipe
1
or upstream pipe of an outlet
2
,
there is a risk of partially filled pipes leading to inaccurate measurements.
•
Installing the device in an upper pipe bend
3
, there is a risk of gas bubbles forming.
Gas bubbles can lead to pressure surges and inaccurate measurement.