OPERATION
5
29
OPTISYS CL 1100
www.krohne.com
03/2016 - 4002492302 - MA OPTISYS CL 1100 R03 en
Step 1: activating the manual hold function
You have activated the manual hold function. To perform the next step and prepare the
calibration procedure you have to return to the measuring mode.
Press
^
^
^
^
until you reach the measuring mode again.
Step 2: preparing the calibration procedure
•
If you calibrate a new sensor, make sure that the sensor is correctly connected to the signal
converter.
•
Check the sensor for damages or dirt deposits.
•
During the calibration procedure, you will have to take a sample. Provide a suitable sample
vessel.
Press >>>> for more than 2.5 seconds, then release the button.
You are on the main menu level. In the upper line on the right side of the display "A" appears, beneath the
main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "quick
setup" and "A1" appears, beneath the submenu language
language
language
language is highlighted.
Press
or
until the submenu manual hold
manual hold
manual hold
manual hold is highlighted.
Press >
>>> to enter the chosen menu.
You are on the parameter level. The option off
off
off
off is highlighted.
Press
or
to choose the option on
on
on
on.
Press
^
^
^
^
to confirm your selection.