4
ELECTRICAL CONNECTIONS
42
OPTISYS SLM 2100
www.krohne.com
01/2017 - 4002737304 - MA OPTISYS SLM 2100 R04 en
The pump connector should be only connected to the OTPISYS SLM 2100 if the external pump is
used as cleaning medium. The terminal connection is outlined in the following diagram.
For installation of the relays follow the steps below:
•
Move the sensor by using the manual operation in the home position and turn off the power of
the instrument.
•
Open the cable drum and electronics compartment doors.
•
Push the prepared cables through the cable feed through and route them to the electronic
compartment.
•
Connect the positive and negative lead according to the connection diagrams above.
•
Connect the shield to one side only e.g. on PCS (process control system) side.
•
Tighten the screw connection of the cable entries securely.
•
Seal all cable entries that are not needed with a plug.
•
Close both compartment doors.
4.11 Protection category
IP 68 (sensor), IP 54 (electronic compartment), IP 44 (cable drum compartment).
Figure 4-9: Connection diagram pump (K3)
1
External voltage
2
PTC resistance (I
max
= 0.3 A)
DANGER!
Do not install the sludge blanket meter in hazardous areas, it can ignite explosive gases!
Summary of Contents for OPTISYS SLM 2100
Page 85: ...NOTES 8 85 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Page 86: ...8 NOTES 86 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Page 87: ...NOTES 8 87 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...