38
www.krohne.com
03/2012 - 4001904901 - HB OPTIWAVE 5200 R01 en
4
ELECTRICAL CONNECTIONS
OPTIWAVE 5200 C/F
4.4 Protection category
INFORMATION!
The device fulfills all requirements per protection category IP 66/67 (equivalent to NEMA type 4X
(housing) and type 6P (antenna)).
DANGER!
Make sure that the cable gland is watertight.
Figure 4-13: How to make the installation agree with protection category IP 67
•
Make sure that the gaskets are not damaged.
•
Make sure that the electrical cables are not damaged.
•
Make sure that the electrical cables agree with the national electrical code.
•
The cables are in a loop in front of the device
1
so water does not go into the housing.
•
Tighten the cable feedthroughs
2
.
•
Close unused cable feedthroughs with dummy plugs
3
.
Summary of Contents for OPTIWAVE 5200 C/F