INSTALLATION
3
25
OPTIWAVE 7300 C
www.krohne.com
04/2010 - 4000172403 - HB OPTIWAVE 7300 R05 en
Bypass chambers
Installation next to tanks containing one liquid and foam
•
The top process connection of the bypass chamber must be above the maximum level of
liquid.
•
The bottom process connection of the bypass chamber must be below the lowest measured
level of liquid.
Figure 3-11: Horizontal cylindrical tanks
1
The device is installed without a stilling well. There are multiple reflections. Refer to the CAUTION! that follows.
2
The device is installed in a stilling well and measures correctly.
CAUTION!
If the device is installed in horizontal cylindrical tank that contains a high dielectric constant
liquid without a stilling well, do not put it on the tank centerline. This will cause multiple
reflections and the device will not measure accurately. Use the Multiple Reflections
Multiple Reflections
Multiple Reflections
Multiple Reflections function in
Supervisor > Advanced Setup > Installation Setup
Supervisor > Advanced Setup > Installation Setup
Supervisor > Advanced Setup > Installation Setup
Supervisor > Advanced Setup > Installation Setup to keep the effects of multiple reflections to a
minimum. For more data, refer to Function description on page 56
(C. Advanced Setup).
HB_OPTIWAVE7300_en_100430_4000172403_R05_D6B.book Page 25 Thursday, May 6, 2010 6:14 PM