3
INSTALLATION
30
OPTIWAVE 7300 C
www.krohne.com
04/2010 - 4000172403 - HB OPTIWAVE 7300 R05 en
If the process connection of the device is smaller than the process connection on the
tank:
•
Make sure the tank connection is level.
•
Use a plate with a slot or a different applicable procedure to adapt the device to the tank.
•
Align the gasket correctly.
•
Lower the antenna carefully into the tank.
•
If necessary, turn the threaded connection on the housing to attach the device to the plate.
•
Tighten the connection.
3.7.3 How to install a device with a hygienic connection
BioControl
®
Equipment needed:
•
Device
•
Gasket
•
Flange bolts (not supplied)
•
Wrench (not supplied)
How to attach a device with a Biocontrol
®
connection
•
Make sure the flange on the nozzle is level.
•
Make sure that you use the applicable gasket for the flange dimensions and the process.
•
Align the gasket correctly on the flange facing of the nozzle.
•
Lower the antenna carefully into the tank.
•
Tighten the flange bolts.
i
Refer to local rules and regulations for the correct torque to apply to the bolts.
INFORMATION!
To make the cleaning of the antenna easier, attach the device to a short socket.
Figure 3-17: BioControl
®
connection
1
BioControl
®
connection on the tank
2
Flange bolts
HB_OPTIWAVE7300_en_100430_4000172403_R05_D6B.book Page 30 Thursday, May 6, 2010 6:14 PM