3.1 Installation
Mounting of the sensor
•
Ensure that you use a gasket suitable for the dimensions of the process connection and the
requirements of the measuring point.
•
Carefully lower the sensor into the pipe.
•
Observe all the relevant regulations.
3.2 Specific conditions of use
The specific conditions of use mentioned in the certificate are as follows:
•
For temperature data, refer to section 2.6 of this supplementary manual.
•
The Conductivity Sensor shall be installed in such a way that danger of electrostatic
discharge is prevented. The equipment may only be cleaned with a wet cloth.
DANGER!
The manufacturer is not liable for any damage or injuries resulting from improper use or use
other than the intended purpose. This applies in particular to hazards due to insufficient
corrosion resistance and suitability of the materials in contact with product.
DANGER!
Installation and setup must be carried out according to the applicable installation standards (e.g.
IEC 60079-14) by qualified personnel trained in explosion protection. Observe the information
contained in the manuals and the supplementary instructions. The installation must always
comply with the following requirements:
•
No external forces are affecting the sensor.
•
The device is accessible for any necessary visual inspections and can be viewed from all
sides.
•
The nameplate is clearly visible.
•
All gaskets are suitable for the intended use.