6
OPERATION
50
WATERFLUX 3070
www.krohne.com
06/2016 - 4004999701 - MA WATERFLUX 3070 - R01 en
6.3.4 Automatic Meter Reading mode (AMR)
The AMR mode provides an option to read the display as base for the pulse output (e.g. use
certain digits of the counter value).
In Automatic Meter Reading (AMR) mode the following settings apply:
•
Menu number 10 (Counter: unit of volume) cannot be set to litre or acre-inch.
•
Menu number 15 (Flow rate: unit of volume / time) cannot be set to litre per second or
acre-inch per day.
6.3.5 Counters and flow rate
The following preferences are available:
•
Unit of volume or time
•
Accuracy of the counters
•
Multiplier value
•
Availability of forward and reverse counter and flow rate on the display
The counter values can be programmed to display a higher accuracy with a decimal dot (menu
number 11). The multiplier value enables the display of volumes with more than 8 digits (menu
number 12).
If a multiplier is selected, the multiplier value is shown in the top right corner of the display.
The value on the display must be multiplied with a factor 10, 100 or 1000 to get the actual counter
value.The decimal dot and multiplier value are not available for the flow rate.
The default setting for number of decimals and the multiplier value is 99. The number of
decimals and multiplier are set automatically by the software, based on the following criteria:
No.
Function
Options
Description
2
AMR mode
0 = Off (default)
Automatic Meter Reading mode (AMR)
1 = On
3
AMR digits
8
Use all digits 87654321
7 -
Use 7 most significant digits 8765432-
- 7
Use 7 least significant digits -7654321
6 - -
Use 6 most significant digits 876543--
- 6 -
-765432-
- - 6
--654321
5 - - -
87654---
- 5 - -
-76543--
- - 5 -
--65432-
- - - 5
---54321
4 - - - -
8765----
- 4 - - -
-7654---
- - 4 - -
--6543--
- - - 4 -
---5432-
- - - - 4
----4321