Software-controlled single-Dosing unit DS3210
page 59
9.2
Fitting the syringe
Two types of syringe can be used for the dosing modules:
Glass syringe SY20
Disposable plastic syringe SY3601
In this chapter, the fitting of the syringe is described based on a glass syringe.
The inner walls of the disposable syringes may be wetted with silicone oil as
lubricant.
Bleeding the syringe
The syringe can be easily filled in the fitted state. To do this, it must first be filled
and bled so that there is no air in the syringe when it is filled later.
1. Fit a needle to the syringe.
2. Fill the syringe with the test liquid.
3. Hold the syringe upright and force residual air out of the syringe and needle.
Fixing the syringe
1. Loosen the fixing screw on the syringe unit slightly (do not remove it!).