page 70
Environmental chambers TC30 and TC40
11.4
TC30/40: Mounting on the DSA100
!
If an HC10 humidity chamber is to be installed it must first be mounted on the
environmental chamber.
1 Raise the assembled dosing unit. Lower the sample table.
2. Unscrew the two hexagonal socket screws underneath the sample table support
using an Allen wrench.
Remove the sample table adapter.
3. Place the chamber on the sample table. Screw the chamber securely to the
sample table using the screws which were removed in Step 2.
11.5
Connectors and connections
11.5.1
Connecting the thermostat
Both environmental chambers are connected to a thermostat. In most cases,
water is used as the thermostat liquid.
!
Do not use silicone oil because it attacks the seals and tubes. Vegetable oils
and ethylene glycol on the other hand are suitable liquids.
Sample table adapter