Operating
instructions
Make
sure
that
the
clamp
properly
encircles
the
conductor,
and
that
there
is
no
air
gap
between
the
jaws.
Also
make
sure
that
you
are
enclosing
all
relevant
conductors
(L1,
L2,
L3
plus
neutral
wire
N
for
3
‐
phase
systems
and
L1
and
N
for
single
phase
systems.)
NOTE
:
Leakage
current
can
be
an
indicator
of
the
effectiveness
of
insulation
on
conductors.
High
levels
of
leakage
current
may
be
present
in
circuits
where
electronic
equipment
with
filters
is
used,
and
can
cause
voltages
that
disrupt
normal
operation
of
equipment.
It
is
possible
to
locate
the
source
of
leakage
current
by
using
a
low
current
leakage
current
clamp
to
take
methodical
measurements
as
described
below.
Measurement
of
Leaking
Current
to
Ground
1.
Set
the
rotary
switch
always
to
the
highest
available
range.
2.
Open
the
jaws
and
insert
the
ground
conductor
through
the
jaws.
Make
sure
the
jaws
are
perpendicular
to
the
conductor.
3.
Close
the
clamp
and
make
sure
the
tops
of
the
jaws
are
properly
shut.
4.
Once
the
reading
stabilizes,
read
the
value.
Use
the
DATA
HOLD
function
if
the
reading
is
difficult
to
read.
M
~
Summary of Contents for PANCONTROL PAN Leckstromzange
Page 1: ...MANUAL PAN Leckstromzange ...
Page 3: ...BEDIENUNGSANLEITUNG PAN Leckstromzange ...
Page 17: ...OPERATING MANUAL PAN Leckstromzange ...
Page 33: ...MANUEL D UTILISATION PAN Leckstromzange Pince de courant de fuite ...
Page 48: ...ISTRUZIONI PER L USO PAN Leckstromzange Pinza per correnti di fuga ...
Page 62: ...GEBRUIKSAANWIJZING PAN Leckstromzange Lekstroomtang ...
Page 76: ...BRUKSANVISNING PAN Leckstromzange strömtång för läckström ...