Figure 8: Purge Block In Scanner
Step 2:
Remove purge block with gloved hand. Clean block with propanol or similar cleaner. If any
holes in the block appear to be clogged they can be cleaned out with either an ultrasonic cleaner or
very fine wire. After cleaning allow block to dry thoroughly.
Step 3:
Spray sides and top with silicone lubricant. Allow lubricant to dry.
Step 4:
Replace block in scanner in correct orientation (See Figure 8). Replace top plate and
modules as in sections 5.1.1 and 5.1.2.
5.3
Training
Please contact your sales representative is further training for maintenance personnel is desired.
Kulite offers training on replacing various scanner components and tubing.