Bit(s)
Field
Value
Description
15
Status A
0
Signifies Status Word A
14-10 Reserved
1111 Reserved
9
Module error
0
All modules are operating normally
1
Communication error with one or more modules
8
Thermostat
0
Scanner is below the thermostat temperature
Scanner is at or above the thermostat temperature
NOTE: This bit is always 0 on scanners without a
heater
1
7
Module D
0
Module D temperature is within the compensated
range
compensation range
1
Module D temperature is outside the compensated
range
6
Module C
0
Module C temperature is within the compensated
range
compensation range
1
Module C temperature is outside the compensated
range
5
Module B
0
Module B temperature is within the compensated
range
compensation range
1
Module B temperature is outside the compensated
range
4
Module A
0
Module A temperature is within the compensated
range
compensation range
1
Module A temperature is outside the compensated
range
3
Scanner
0
Scanner is at or below the maximum allowed
operating temperature
operating range
1
Scanner is above the maximum allowed operating
temperature
2
Purge State
0
Purge motor is operating normally
1
Purge motor is stuck
NOTE: This bit is always 0 on units without a
motorized purge
1-0
Purge Position
00
Purge is fully closed
01
Purge is fully opened
10
Purge is closing
11
Purge is opening
NOTE: These bits are always 00 on units without a
motorized purge
NOTE: These bits are undefined if the purge is stuck
Table 12: KMPS Status Word A