KMPS-1-64 Manual
12/18
KM 8000 Rev. A
Page 38 of 40
Copyright © 2018 Kulite Semiconductor Products, Inc.
All Rights Reserved.
40966
03C5
Max compensated
temperature - module B
1
int16
Returns 0 if module missing
40967
03C6
Max compensated
temperature - module C
1
int16
Returns 0 if module missing
40968
03C7
Max compensated
temperature - module D
1
int16
Returns 0 if module missing
40969
03C8
Max operating
temperature
1
int16
40970
03C9
Purge position
1
uint16
0-100
0 indicates fully open
100 indicates fully closed
[In unitswithout a purge, reads
always return 0]
40971
03CA
KMPS reference
temperature
1
int16
Temperature of reference channel
used for thermal control
40972
03CB
KMPS Status A
1
unit16
See Error! Reference source not
found.
40973
03CC
KMPS Status B
1
uint16
Table 14 MODBUS/TCP Holding Registers
Ref
[Dec]
Address
[Hex]
Description
Length(words
)
Format
Range/Comments
30001
0000
Pressure - channel 0
2
IEEE-754 (big endian)
Returns NaN if module missing
30003
0002
Pressure - channel 1
2
IEEE-754 (big endian)
Returns NaN if module missing
30005
0004
Pressure - channel 2
2
IEEE-754 (big endian)
Returns NaN if module missing
.
.
.
30123
007A
Pressure - channel 61
2
IEEE-754 (big endian)
Returns NaN if module missing
30125
007C
Pressure - channel 62
2
IEEE-754 (big endian)
Returns NaN if module missing
30127
007E
Pressure - channel 63
2
IEEE-754 (big endian)
Returns NaN if module missing
30129
0080
Temperature -channel 0
2
IEEE-754 (big endian)
Returns NaN if module missing
30131
0082
Temperature -channel 1
2
IEEE-754 (big endian)
Returns NaN if module missing
30133
0084
Temperature -channel 2
2
IEEE-754 (big endian)
Returns NaN if module missing
.
.
.
30251
00FA
Temperature - channel 61
2
IEEE-754 (big endian)
Returns NaN if module missing
30253
00FC
Temperature - channel 62
2
IEEE-754 (big endian)
Returns NaN if module missing
30255
00FE
Temperature - channel 63
2
IEEE-754 (big endian)
Returns NaN if module missing
30257
0100
Full-scale pressure - chan
0
2
IEEE-754 (big endian)
Returns NaN if module missing
30259
0102
Full-scale pressure - chan
1
2
IEEE-754 (big endian)
Returns NaN if module missing
30261
0104
Full-scale pressure - chan
2
2
IEEE-754 (big endian)
Returns NaN if module missing
.
.
.
30379
017A
Full-scale pressure - chan
61
2
IEEE-754 (big endian)
Returns NaN if module missing
30381
017C
Full-scale pressure - chan
62
2
IEEE-754 (big endian)
Returns NaN if module missing
30383
017E
Full-scale pressure - chan
63
2
IEEE-754 (big endian)
Returns NaN if module missing
30385
0180
Pressure type - channel 0
1
uint16
1 - Absolute 2 - Gauge
Summary of Contents for KMPS-1-64 Series
Page 1: ......