KMPS-1-64 Manual
12/18
KM 8000 Rev. A
Page 39 of 40
Copyright © 2018 Kulite Semiconductor Products, Inc.
All Rights Reserved.
30386
0181
Pressure type - channel 1
1
uint16
1 - Absolute 2 - Gauge
30387
0182
Pressure type - channel 2
1
uint16
1 - Absolute 2 - Gauge
.
.
.
30446
01BD
Pressure Type - channel
61
1
uint16
1 - Absolute 2 - Gauge
30447
01BE
Pressure Type - channel
62
1
uint16
1 - Absolute 2 - Gauge
30448
01BF
Pressure Type - channel
63
1
uint16
1 - Absolute 2 - Gauge
30449
01C0
Min compensated
temperature - module A
1
int16
Returns 0 if module missing
30450
01C1
Min compensated
temperature - module B
1
int16
Returns 0 if module missing
30451
01C2
Min compensated
temperature - module C
1
int16
Returns 0 if module missing
30452
01C3
Min compensated
temperature - module D
1
int16
Returns 0 if module missing
30453
01C4
Max compensated
temperature - module A
1
int16
Returns 0 if module missing
30454
01C5
Max compensated
temperature - module B
1
int16
Returns 0 if module missing
30455
01C6
Max compensated
temperature - module C
1
int16
Returns 0 if module missing
30456
01C7
Max compensated
temperature - module D
1
int16
Returns 0 if module missing
30457
01C8
Max operating
temperature
1
int16
30458
01C9
Purge position
1
uint16
0-100
0 indicates fully open
100 indicates fully closed
[In units without a purge, reads
always return 0]
30459
01CA
KMPS reference
temperature
1
int16
Temperature of reference channel
used for thermal control
30460
01CB
KMPS Status A
1
unit16
See Error! Reference source not
found.
30461
01CC
KMPS Status B
1
uint16
Table 15 MODBUS/TCP Input Registers