- 15 -
GB
DE
GB
FR
ES
IT
PT
NL
SE
DK
NO
FI
GR
RU
Exposure 10 seconds polymerisation
Briefly press the “
STAND-BY
” button
. The light emission will turn on.
The 10 second LED is on showing that this program is running.
Once the curing cycle is complete, the unit will automatically turn the light off.
Acoustical signal:
The following acoustical signals (beep) are emitted:
1 beep
while press the “
STAND-BY
” button.
1 beep
at the end of the exposure time.
Exposure 20 seconds “Soft-Start” polymerisation
Press and keep pressed for at least one second the “
STAND-BY
” button
. The light emission will turn on.
The 20 second LED is on showing that this program is running.
Once the curing cycle is complete, the unit will automatically turn the light off and the unit returns to the 10 second
mode (10 second LED on).
Acoustical signal:
The following acoustical signals (beep) are emitted:
1 beep
while press the “
STAND-BY
” button.
1 beep
after 1 second for acoustic indication of “
Soft-Start
”-mode activation.
1 beep
after 10 seconds of exposure time.
1 beep
at the end of the exposure time.
NOTE
The exposure cycle can be interrupted at any time, before the exposure time is over,
by pressing again the “
STAND-BY
” button.
Acoustical signal:
An acoustical signal (
beep
) is emitted, when the “
STAND-BY
” button is pressed to interrupt the cycle.
6.2
Measurement of the Light Intensity
The charging unit of the Translux Wave is equipped with an integrated light intensity meter (
8
).
CAUTION
Before each treatment always check the light intensity by using the light intensity meter
integrated in the Translux Wave charging unit.
• Without applying pressure, place the light guide flat and directly on the testing area (
8
).
Activate the unit by pressing the “
STAND-BY
” button
(whichever exposure time selected).
The light intensity measuring is immediately indicated by the 2-colour LED (Symbol
) placed at the bottom right
of the testing area.
GREEN LED
: The light intensity is sufficient for the treatment.
YELLOW LED
: The light intensity is insufficient and patient/s must not be treated.
CAUTION
If the light intensity is insufficient, please carry out the following controls and measures:
– The light guide has not been inserted into the hand piece (see paragraph 5.4 Inserting the fibre optic
into the hand piece).
– Check the light guide for contamination or defects.
– Clean the light guide of any contamination (see paragraph 8 Cleaning, disinfection and sterilisation)
or replace the damaged light guide with a new one.
Use original light guides only.
Kulzer GmbH will provide only original light guides supplied by Mectron S.p.A.
WARNING
If the measures described above do not result in any improvement, do not use the device for any treatment.
Switch off the unit (disconnect the charging unit from the electrical power source) and ensure that
unauthorised unit re-activation is not possible. Contact a Kulzer and/or authorized Kulzer service partners
only (see paragraph 14 Service).
2x
Summary of Contents for 05100081
Page 2: ...2 4 1 4 2 3 5 2 6c 6b 6a 5 4 3 6 9 10 7 1 8...
Page 3: ...4 2 1 4 2 2 1 5 4 5 4 X X 3 2 2 1 3...
Page 4: ...5 5 6 2 5 4 2 2 8...
Page 37: ...33 DE...
Page 38: ...34 DE nderungen vorbehalten Stand 2021 08...
Page 71: ...33 GB...
Page 72: ...34 GB Subject to modifications Dated 2021 08...
Page 105: ...33 FR...
Page 106: ...34 FR Sujet modifications Mise jour de l information 2021 08...
Page 139: ...33 ES...
Page 140: ...34 ES Sujeto a modificaciones Revisi n 2021 08...
Page 173: ...33 IT...
Page 174: ...34 IT Soggetto a modifiche Aggiornamento al 2021 08...
Page 207: ...33 NL...
Page 208: ...34 NL De informatie in dit document kan gewijzigd worden Status 2021 08...
Page 241: ...33 SE...
Page 242: ...34 SE Kan komma att ndras Version 2021 08...
Page 275: ...33 FI...
Page 276: ...34 FI Muutosten kohteena Painos 2021 08...
Page 282: ...6 GR 3 1 3 3 1 1 3 3 1 4 3 2 3 2 1 3 Mectron S p A Translux Wave Translux Wave...
Page 293: ...17 GR 7 5 Translux Wave 2 2 8 8 1 8 2 1 2 pH pH 7 8 2 1 8 3 pH pH 7...
Page 295: ...19 GR 8 5 pH pH7 8 6 pH pH7 2x...
Page 296: ...20 GR 8 6 1 8 6 2 Kulzer Kulzer 1 2 3 4 5 2 12 6 7 9 5 2x 2x...
Page 297: ...21 GR 10 Kulzer WEEE ElektroG...
Page 309: ...33 GR...
Page 310: ...34 GR 2021 08...
Page 311: ......
Page 312: ...5 3 4 2 3 8 5 8 6 1 4 1 2 3 8 6 2 11 12 12...
Page 313: ......