8.3.1 PROTECTIVE LIGHT CONE – Inspection before the sterilisation
Before proceeding with the sterilisation process, carefully check the protective light cone for damaged surfaces,
veins discolouration and contamination.
WARNING
Do not
use a damaged protective light cone. If damage is apparent, replace the protective light cone with
a new one.
A damaged protective light cone cannot provide enough protection from the unit light.
If the protective light cone is still contaminated, repeat the cleaning and disinfection.
8.4
LIGHT GUIDE and PROTECTIVE LIGHT CONE – Packaging before the sterilisation
The light guide and the protective light cone can be sterilised by using a single-use medical steam sterilisation pouch
of the appropriate size.
CAUTION
The light guide and the protective light cone must be packed individually in a single pouch.
When sterilising multiple instruments in one autoclave, ensure that the autoclave maximum load
is not exceeded.
WARNING
Ensure that the inner pouch is large enough to contain the single instruments, without stressing the seals
or tearing the packaging.
8.4.1 LIGHT GUIDE and PROTECTIVE LIGHT CONE – Sterilisation
8.4.1.1 Sterilisation method
• Carry out the sterilisation by means of a pre-vacuum steam autoclaving.
CAUTION
Never use
any other sterilisation method, because of potential incompatibility with the materials with which
is made the light guide.
DO NOT USE the following sterilisation methods:
Ethylene oxide sterilisation, hot air sterilisation,
flash autoclaving, STERRAD sterilisation, STERIS system or comparable sterilisation systems.
DO NOT sterilise the light guide and the protective light cone by using:
Hydrogen peroxide, peracetic acid
system, formaldehyde and glutaraldehyde sterilisation.
8.4.1.2 Sterilisation parameters
•
Minimum sterilisation parameters that provide a 10
–6
sterility assurance level – (SAL).
Cycle Type
Minimum Temperature
Minimum Exposure Time (pouch)
Minimum Dry Time
3 times Pre-vacuum
(min. pressure 60 mBar)
132°C + 3°C
(270°F + 5°F)
4 minutes
20 minutes
CAUTION
Do not exceed 135°C (275°F).
Summary of Contents for 05100081
Page 2: ...2 4 1 4 2 3 5 2 6c 6b 6a 5 4 3 6 9 10 7 1 8...
Page 3: ...4 2 1 4 2 2 1 5 4 5 4 X X 3 2 2 1 3...
Page 4: ...5 5 6 2 5 4 2 2 8...
Page 37: ...33 DE...
Page 38: ...34 DE nderungen vorbehalten Stand 2021 08...
Page 71: ...33 GB...
Page 72: ...34 GB Subject to modifications Dated 2021 08...
Page 105: ...33 FR...
Page 106: ...34 FR Sujet modifications Mise jour de l information 2021 08...
Page 139: ...33 ES...
Page 140: ...34 ES Sujeto a modificaciones Revisi n 2021 08...
Page 173: ...33 IT...
Page 174: ...34 IT Soggetto a modifiche Aggiornamento al 2021 08...
Page 207: ...33 NL...
Page 208: ...34 NL De informatie in dit document kan gewijzigd worden Status 2021 08...
Page 241: ...33 SE...
Page 242: ...34 SE Kan komma att ndras Version 2021 08...
Page 275: ...33 FI...
Page 276: ...34 FI Muutosten kohteena Painos 2021 08...
Page 282: ...6 GR 3 1 3 3 1 1 3 3 1 4 3 2 3 2 1 3 Mectron S p A Translux Wave Translux Wave...
Page 293: ...17 GR 7 5 Translux Wave 2 2 8 8 1 8 2 1 2 pH pH 7 8 2 1 8 3 pH pH 7...
Page 295: ...19 GR 8 5 pH pH7 8 6 pH pH7 2x...
Page 296: ...20 GR 8 6 1 8 6 2 Kulzer Kulzer 1 2 3 4 5 2 12 6 7 9 5 2x 2x...
Page 297: ...21 GR 10 Kulzer WEEE ElektroG...
Page 309: ...33 GR...
Page 310: ...34 GR 2021 08...
Page 311: ......
Page 312: ...5 3 4 2 3 8 5 8 6 1 4 1 2 3 8 6 2 11 12 12...
Page 313: ......