- 27 -
GB
DE
GB
FR
ES
IT
PT
NL
SE
DK
NO
FI
GR
RU
Guidance and manufacturer’s declaration – Electromagnetic immunity
Input a.c. power port
The Translux Wave device is designed to work in the electromagnetic environment specified below.
The client or user of the Translux Wave device should always make sure that it is used only in such environment.
Phenomenon
Basic EMC standard
or test method
Immunity test levels
Electromagnetic environment – Guidance
Electrical fast
transients /
bursts
a) l) o)
IEC 61000-4-4
±2 kV contact
100 KHz repetition frequency
The quality of the power supply should be
that of a typical commercial or hospital
environment.
Surges
Line-to-line
a) b) j) o)
IEC 61000-4-5
± 0,5 kV, ± 1 kV
The quality of the power supply should be
that of a typical commercial or hospital
environment.
Surges
Line-to-ground
a) b) j) k) o)
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ± 2kV
The quality of the power supply should be
that of a typical commercial or hospital
environment.
Conducted
disturbances induce
by RF fields
c) d) o)
IEC 61000-4-6
3 V
m)
0,15 MHz - 80 MHz
6 V
m)
in ISM bands between
0,15 MHz and 80 MHz
n)
80 % AM at 1 KHz
e)
Portable or mobile RF communication
device must not be used in close proximity
of the product, including its cables, except
when these respect the distances of sepa-
ration recommended and calculated from
the equation applicable to the frequency of
the transmitter.
Voltage dips
f) p) r)
IEC 61000-4-11
0% U
T
; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
The quality of the power supply should be
that of a typical commercial or hospital
environment.
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
h)
Single phase: at 0°
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
0 % U
T
; 250/300 cycle
h)
The quality of the power supply should be
that of a typical commercial or hospital
environment.
a) The test may be performed at any one power input voltage within the Translux Wave RATED voltage range. If the Translux Wave is tested at one power input
voltage, it is not necessary to re-test at additional voltages.
b) All Translux Wave cables are attached during the test.
c) Calibration for current injection clamps shall be performed in a 150 Ω system.
d) If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency shall be used in the ISM or amateur radio band. This
applies to each ISM and amateur radio band within the specified frequency range.
e) Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.
f) Translux Wave with a d.c. power input intended for use with a.c.-to-d.c. converters shall be tested using a converter that meets the specifications of the MANU-
FACTURER of the Translux Wave. The IMMUNITY TEST LEVELS are applied to the a.c. power input of the converter .
g) Applicable only to Translux Wave connected to single-phase a.c. mains.
h) E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
i) Translux Wave with RATED input current greater than 16 A / phase shall be interrupted once for 250/300 cycles at any angle and at all phases at the same time
(if applicable). Translux Wave with battery backup shall resume line power operation after the test. For Translux Wave with RATED input current not exceeding
16 A, all phases shall be interrupted simultaneously.
j) Translux Wave that do not have a surge protection device in the primary power circuit may be tested only at ± 2 kV line(s) to earth and ± 1 kV line(s) to line(s).
k) Not applicable to CLASS II Translux Wave.
l) Direct coupling shall be used.
m) r.m.s. , before modulation is applied.
n) The ISM (industrial scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to
27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0, 15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz,
28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
o) Applicable to Translux Wave with RATED input current less than or equal to 16 A / phase and Translux Wave with RATED input current greater than 16 A / phase.
p) Applicable to Translux Wave with RATED input current less than or equal to 16 A / phase.
q) At some phase angles, applying this test to Translux Wave with transformer mains power input might cause an overcurrent protection device to open. This can
occur due to magnetic flux saturation of the transformer core after the voltage dip. If this occurs, the Translux Wave shall provide BASIC SAFETY during and
after the test.
r) For Translux Wave that have multiple voltage settings or auto ranging voltage capability, the test shall be performed at the minimum and maximum RATED input
voltage. Translux Wave with a RATED input voltage range of less than 25 % of the highest RATED input voltage shall be tested at one RATED input voltage within
the range.
Summary of Contents for 05100081
Page 2: ...2 4 1 4 2 3 5 2 6c 6b 6a 5 4 3 6 9 10 7 1 8...
Page 3: ...4 2 1 4 2 2 1 5 4 5 4 X X 3 2 2 1 3...
Page 4: ...5 5 6 2 5 4 2 2 8...
Page 37: ...33 DE...
Page 38: ...34 DE nderungen vorbehalten Stand 2021 08...
Page 71: ...33 GB...
Page 72: ...34 GB Subject to modifications Dated 2021 08...
Page 105: ...33 FR...
Page 106: ...34 FR Sujet modifications Mise jour de l information 2021 08...
Page 139: ...33 ES...
Page 140: ...34 ES Sujeto a modificaciones Revisi n 2021 08...
Page 173: ...33 IT...
Page 174: ...34 IT Soggetto a modifiche Aggiornamento al 2021 08...
Page 207: ...33 NL...
Page 208: ...34 NL De informatie in dit document kan gewijzigd worden Status 2021 08...
Page 241: ...33 SE...
Page 242: ...34 SE Kan komma att ndras Version 2021 08...
Page 275: ...33 FI...
Page 276: ...34 FI Muutosten kohteena Painos 2021 08...
Page 282: ...6 GR 3 1 3 3 1 1 3 3 1 4 3 2 3 2 1 3 Mectron S p A Translux Wave Translux Wave...
Page 293: ...17 GR 7 5 Translux Wave 2 2 8 8 1 8 2 1 2 pH pH 7 8 2 1 8 3 pH pH 7...
Page 295: ...19 GR 8 5 pH pH7 8 6 pH pH7 2x...
Page 296: ...20 GR 8 6 1 8 6 2 Kulzer Kulzer 1 2 3 4 5 2 12 6 7 9 5 2x 2x...
Page 297: ...21 GR 10 Kulzer WEEE ElektroG...
Page 309: ...33 GR...
Page 310: ...34 GR 2021 08...
Page 311: ......
Page 312: ...5 3 4 2 3 8 5 8 6 1 4 1 2 3 8 6 2 11 12 12...
Page 313: ......