Test specifications for enclosure port immunity to RF wireless communications equipment
The Translux Wave device is designed to work in an electromagnetic environment in which RF radiated disturbances are
kept under control. The purchaser or user of the Translux Wave device can contribute to prevent electromagnetic inter-
ference by ensuring a minimum distance between the RF mobile and cordless communication device (transmitters) and
the Translux Wave device, as recommended below, in relation to the maximum output power of the radio-communication
devices.
Test frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance (m)
Immunity test
level (V/m)
385
380 – 390
TETRA 400
Pulse modulation
b)
18 Hz
1,8
0,3
27
450
430 – 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band 13,
17
Pulse modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
b)
18 Hz
2
0,3
28
870
930
1720
1700 – 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse modulation
b)
217 Hz
2
0,3
28
1845
1970
2450
2400 – 2570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse modulation
b)
217 Hz
2
0,3
28
5420
5100 – 5800
WLAN 802.11
a/n
Pulse modulation
b)
217 Hz
0,2
0,3
9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
Translux Wave may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the Translux Wave device, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Summary of Contents for 05100081
Page 2: ...2 4 1 4 2 3 5 2 6c 6b 6a 5 4 3 6 9 10 7 1 8...
Page 3: ...4 2 1 4 2 2 1 5 4 5 4 X X 3 2 2 1 3...
Page 4: ...5 5 6 2 5 4 2 2 8...
Page 37: ...33 DE...
Page 38: ...34 DE nderungen vorbehalten Stand 2021 08...
Page 71: ...33 GB...
Page 72: ...34 GB Subject to modifications Dated 2021 08...
Page 105: ...33 FR...
Page 106: ...34 FR Sujet modifications Mise jour de l information 2021 08...
Page 139: ...33 ES...
Page 140: ...34 ES Sujeto a modificaciones Revisi n 2021 08...
Page 173: ...33 IT...
Page 174: ...34 IT Soggetto a modifiche Aggiornamento al 2021 08...
Page 207: ...33 NL...
Page 208: ...34 NL De informatie in dit document kan gewijzigd worden Status 2021 08...
Page 241: ...33 SE...
Page 242: ...34 SE Kan komma att ndras Version 2021 08...
Page 275: ...33 FI...
Page 276: ...34 FI Muutosten kohteena Painos 2021 08...
Page 282: ...6 GR 3 1 3 3 1 1 3 3 1 4 3 2 3 2 1 3 Mectron S p A Translux Wave Translux Wave...
Page 293: ...17 GR 7 5 Translux Wave 2 2 8 8 1 8 2 1 2 pH pH 7 8 2 1 8 3 pH pH 7...
Page 295: ...19 GR 8 5 pH pH7 8 6 pH pH7 2x...
Page 296: ...20 GR 8 6 1 8 6 2 Kulzer Kulzer 1 2 3 4 5 2 12 6 7 9 5 2x 2x...
Page 297: ...21 GR 10 Kulzer WEEE ElektroG...
Page 309: ...33 GR...
Page 310: ...34 GR 2021 08...
Page 311: ......
Page 312: ...5 3 4 2 3 8 5 8 6 1 4 1 2 3 8 6 2 11 12 12...
Page 313: ......