Draining the contaminated water
Close the pressure pot, place the silicone hose on the hose nozzle (8) located on the rear wall of the unit for emptying,
press the mains switch, open the water drain valve (8) and press the water drain program (Settings menu).
After the water has been expelled from the pressure pot, press the button for the water drain program again.
Close the water drain valve, turn off the mains switch.
To empty the water reservoir, use only a silicone hose with a temperature resistance of up to 120°C (248°F).
A suitable hose can be ordered from Kulzer (Order No. 66009729 – silicone hose, red).
The silicone hose
is not included in the scope of delivery!
When emptying the water reservoir, the silicone hose may heat up considerably.
Risk of scalding!
7.5
Putting out of operation
Discharge contaminated water (see chapter 7.4 Draining the contaminated water)
Switch off the unit with the power switch (
5
)
Unplug the unit (
6
)
Any material residues should be removed with a sponge cloth
7.5.1 Disposal of old equipment according to WEEE
Electrical and Electronic Equipment Act (ElektroG)
This Act sets out requirements for electrical and electronic equipment pursuant to directive 2002 / 96 / EG issued by the
European Parliament and the European Council of 2005-05-03. Its main purpose is to prevent waste from electrical and
electronic equipment and to promote reuse, recycling and other forms of recovery to reduce both the volume of waste for
disposal and the inclusion in waste of harmful substances from electrical and electronic equipment.
For detailed information on professional disposal of disused old devices please contact your dealer or Kulzer
subsidiary directly in your country.
Important!
Marked equipment must not be brought to local waste disposal centres!
8 Maintenance
The unit must be unplugged (
6
) before it is cleaned. Use a moist cloth to clean the unit and make sure that no water
penetrates into the unit. Do not use aggressive cleaning agents (scouring powder, solvents).
The unit consists of Makrolon components which may only be cleaned with mild cleaning agents or ethanol (96%).
Spare fuses can be purchased from Kulzer or specialized dealers and may be exchanged by the customer.
Attention!
The unit must be unplugged before exchanging fuses!