Draining the contaminated water

Close the pressure pot, place the silicone hose on the hose nozzle (8) located on the rear wall of the unit for emptying,  
press the mains switch, open the water drain valve (8) and press the water drain program (Settings menu).

After the water has been expelled from the pressure pot, press the button for the water drain program again.  
Close the water drain valve, turn off the mains switch.

 

To empty the water reservoir, use only a silicone hose with a temperature resistance of up to 120°C (248°F).  
A suitable hose can be ordered from Kulzer (Order No. 66009729 – silicone hose, red). 

The silicone hose  

is not included in the scope of delivery!

 

When emptying the water reservoir, the silicone hose may heat up considerably. 

Risk of scalding!

7.5 

Putting out of operation

 

Discharge contaminated water (see chapter 7.4 Draining the contaminated water)

 

Switch off the unit with the power switch (

5

)

 

Unplug the unit (

6

)

 

Any material residues should be removed with a sponge cloth

7.5.1  Disposal of old equipment according to WEEE

Electrical and Electronic Equipment Act (ElektroG)

This Act sets out requirements for electrical and electronic equipment pursuant to directive 2002 / 96 / EG issued by the  
European Parliament and the European Council of 2005-05-03. Its main purpose is to prevent waste from electrical and  
electronic equipment and to promote reuse, recycling and other forms of recovery to reduce both the volume of waste for  
disposal and the inclusion in waste of harmful substances from electrical and electronic equipment.

 

For detailed information on professional disposal of disused old devices please contact your dealer or Kulzer 
 subsidiary directly in your country.

 Important!

  Marked equipment must not be brought to local waste disposal centres!

8 Maintenance

The unit must be unplugged (

6

) before it is cleaned. Use a moist cloth to clean the unit and make sure that no water 

 penetrates into the unit. Do not use aggressive cleaning agents (scouring powder, solvents).

The unit consists of Makrolon components which may only be cleaned with mild cleaning agents or ethanol (96%).  
Spare fuses can be purchased from Kulzer or specialized dealers and may be exchanged by the customer.

 Attention!

  The unit must be unplugged before exchanging fuses!

Summary of Contents for 66057661

Page 1: ...Das ist Zahntechnik DE Betriebsanleitung GB Operating Instruction FR Mode d emploi ES Instrucciones de uso IT Istruzioni per l uso NL Gebruiksaanwijzing Palamat Premium DE GB FR ES IT NL...

Page 2: ...n of the unit 5 6 Setting up installation and putting into operation 6 7 Working with the Palamat Premium 7 7 1 Pressure pot 7 7 2 Filling with water 7 7 2 1 Overtemperature protection 8 7 3 Polymeriz...

Page 3: ...2 Designation and type of the unit Designation of the machine Type of machine Valid from Serial no Polymerization unit for cold and hot curing resins Palamat Premium 2019 04 001 1 3 EC Declaration of...

Page 4: ...ore be taken during transportation as well as storage The equipment shipped by Kulzer was checked thoroughly prior to ship ment The unit is correctly protected and packed when delivered Check the unit...

Page 5: ...which was not carried out by Kulzer service partners or by personnel trained by us or in cases where original spare accessory parts were not used during a part replacement Kulzer GmbH is not liable C...

Page 6: ...se red 66056911 Flask basket 5 Description of the unit Front view 1 Cover 4 Pressure Gauge 5 On Off power switch with LED indicator 2 Pressure Pot 3 TFT Display Supply connections rear 6 Mains connect...

Page 7: ...s shall be excluded in case of malfunctions or damage resulting from inadequate compressed air supply See special accessories chapter 4 Mains connection Prior to operating check whether the nominal vo...

Page 8: ...t be in upright position Turn the lid by 90 tilt it and remove it from the pressure pot After completing the polymerization process the handle must be in a perpendicular position to open the lid The l...

Page 9: ...beep will sound The screen then changes to the Edit display for the respective program Home display In this area the temperature D pressure E on off BAR and time F can be configured individually Pres...

Page 10: ...ightness of the screen does not change J Changing the temperature unit C F 7 3 3 Start pause a process Press one of the buttons 15 16 17 on the Home display to select the corresponding program The cor...

Page 11: ...ding to WEEE Electrical and Electronic Equipment Act ElektroG This Act sets out requirements for electrical and electronic equipment pursuant to directive 2002 96 EG issued by the European Parliament...

Page 12: ...ce partner E 99 Sensor break short circuit water bath or automatic software safety shutdown due to high temperature Press the Home button 11 Turn off the device Check water level see section 7 2 Allow...

Page 13: ...12 GB Subject to modifications Dated 2019 04 12 Document history 2019 04 First edition...

Page 14: ...facturer Kulzer GmbH Leipziger Stra e 2 63450 Hanau Germany W20960 99001462 11 2019 04 Distributed in USA Canada exclusively by Kulzer LLC 4315 South Lafayette Blvd South Bend IN 46614 2517 1 800 431...

Reviews: