 Attention! 

  Check the content of the shipping carton and compare with the scope of delivery described in the operating  

instructions (see chapter 4 scope of delivery).

 

Prior to operation it must be verified whether the information provided on the type plate correspond with the local 
power supply data.

 

The Palamat Premium polymerization unit features a cold-appliance plug and a safety plug. Power cord and plug 
must be checked for damage prior to the use. If any damage is found, the unit must not be connected to the mains 
supply.

 

The Palamat Premium polymerization must be connected to the mains supply via properly installed sockets with  
protective conductor connections (protection class I). The user must have easy access to the plug and mains socket.

7.1 

Pressure pot

To open the lid (

1

), locking lever must be in upright position. Turn the lid by 90°, tilt it and remove it from the pressure pot.

After completing the polymerization process, the handle must be in a perpendicular position to open the lid.  
The lid can only be removed if it has been lowered into the pot after release of pressure.  
The lid is inserted and closed in reverse order. It must ensured that the supporting strut of the lid fits into the grooves on the 
pressure pot rim.

Opening and closing the pressure pot

1

3

2

1)

  Place handle in perpendicular position and turn lid by 90°

2)

   Tilt lid slightly

3)

   Remove lid

The pot is closed in reverse order.

7.2 

Filling with water

 

The unit must not be heated without water. Minimum filling level: 4 cm. Even if objects are placed in,  
the filling level is always below the upper end of the air supply fitting.  
Do not fill the unit under running water. Kulzer recommends filling the device using a suitable container  
with a spout. 
Prior to filling with water, check if the water discharge valve at the rear of the unit is closed.

 

Use only pure tap water for processing.

 

Distilled, deionized water or other liquids may not be used!

Summary of Contents for 66057661

Page 1: ...Das ist Zahntechnik DE Betriebsanleitung GB Operating Instruction FR Mode d emploi ES Instrucciones de uso IT Istruzioni per l uso NL Gebruiksaanwijzing Palamat Premium DE GB FR ES IT NL...

Page 2: ...n of the unit 5 6 Setting up installation and putting into operation 6 7 Working with the Palamat Premium 7 7 1 Pressure pot 7 7 2 Filling with water 7 7 2 1 Overtemperature protection 8 7 3 Polymeriz...

Page 3: ...2 Designation and type of the unit Designation of the machine Type of machine Valid from Serial no Polymerization unit for cold and hot curing resins Palamat Premium 2019 04 001 1 3 EC Declaration of...

Page 4: ...ore be taken during transportation as well as storage The equipment shipped by Kulzer was checked thoroughly prior to ship ment The unit is correctly protected and packed when delivered Check the unit...

Page 5: ...which was not carried out by Kulzer service partners or by personnel trained by us or in cases where original spare accessory parts were not used during a part replacement Kulzer GmbH is not liable C...

Page 6: ...se red 66056911 Flask basket 5 Description of the unit Front view 1 Cover 4 Pressure Gauge 5 On Off power switch with LED indicator 2 Pressure Pot 3 TFT Display Supply connections rear 6 Mains connect...

Page 7: ...s shall be excluded in case of malfunctions or damage resulting from inadequate compressed air supply See special accessories chapter 4 Mains connection Prior to operating check whether the nominal vo...

Page 8: ...t be in upright position Turn the lid by 90 tilt it and remove it from the pressure pot After completing the polymerization process the handle must be in a perpendicular position to open the lid The l...

Page 9: ...beep will sound The screen then changes to the Edit display for the respective program Home display In this area the temperature D pressure E on off BAR and time F can be configured individually Pres...

Page 10: ...ightness of the screen does not change J Changing the temperature unit C F 7 3 3 Start pause a process Press one of the buttons 15 16 17 on the Home display to select the corresponding program The cor...

Page 11: ...ding to WEEE Electrical and Electronic Equipment Act ElektroG This Act sets out requirements for electrical and electronic equipment pursuant to directive 2002 96 EG issued by the European Parliament...

Page 12: ...ce partner E 99 Sensor break short circuit water bath or automatic software safety shutdown due to high temperature Press the Home button 11 Turn off the device Check water level see section 7 2 Allow...

Page 13: ...12 GB Subject to modifications Dated 2019 04 12 Document history 2019 04 First edition...

Page 14: ...facturer Kulzer GmbH Leipziger Stra e 2 63450 Hanau Germany W20960 99001462 11 2019 04 Distributed in USA Canada exclusively by Kulzer LLC 4315 South Lafayette Blvd South Bend IN 46614 2517 1 800 431...

Reviews: