7.2 Installation
The HiLite
®
power 3D unit must be placed on a solid, fire-resistant surface in a way to ensure a safe horizontal position
(laboratory desk, base frames). The ambient temperature should not exceed 45°C (113°F).
Place of installation:
Table / desk with a loading capacity of at least 10 kg.
Dimensions of table / desk
(w x h x d): approx. 250 x 250 x 360 mm
CAUTION
The following instructions must be observed:
The rear panel of the device must have a minimum distance of at least 10 cm to each wall in order not to block
the air supply for the built-in fan. There is a risk of overheating if the air required for cooling cannot circulate
freely around the device. The service life or lamp power may be affected.
The unit must be set up in a way to ensure that the air at the bottom can be blown off freely.
7.3
Function check
CAUTION
The device will only function properly if its components are in perfect condition.
Check the device components thoroughly before using the device!
Ensure that:
no device components are damaged,
the components are free of residue and dirt,
defective parts are replaced,
the plug and the insulation of the power cord are not damaged,
the mains plug and the socket must be compatible.
7.4
Mains connection
At the rear of the unit there is a cold appliance connection (socket) with integrated mains fuses (2 pieces) and the voltage selector.
CAUTION
After setting up the unit, free access to the mains plug must be ensured in order
to unplug the unit.
To avoid damage to the unit, the voltage selector needs to be set to the correct
input voltage before connecting the power cord and the commissioning.
Power cord included in the scope of delivery
Use in Europe
Voltage: 230 V
Use in USA / Japan
Voltage USA: 115 V / Voltage Japan: 100 V
ELECTRIC SHOCK
Wet conditions and moisture may cause short circuits. The following must be ensured:
The unit may only be operated in dry rooms / areas.
7.5
Installation filter disc, flash lamp and chip card
A maintenance flap is located on the side of the unit to allow the removal of the flash lamp and the respective chip card
for the lamp. Additionally you can find a small hole for emergency opening of drawer.
To open the maintenance flap, loosen the self captivating plug locking screws, lift the maintenance flap and remove it.
Insert filter disc into notch. Detailed description of emergency opening procedure you can find in paragraph 10.5.