CAUTION
While working with light reflectors with covered holes heat will accumulate in the unit.
Polymerisation processes will not be reproducible and the unit may switch off owing to excess temperature.
To restart the unit, let it cool down for approx. 30 minutes.
Observe the following information:
Insert only clean light reflectors.
Holes may only be covered to the extent required for positioning the prosthetic restoration.
CAUTION
The filter disc ensures that the material is not exposed to any undesired radiation and reduces the generation of heat.
Observe the following information:
The unit may only be operated with filter disc.
Clean the filter disc within regular intervals.
CAUTION
The filter collects dust and contamination from the laboratory and avoids the output of light from the bottom
of the unit. Moreover the fireproof filter protects the unit if it is heated excessively.
Observe the following information:
The unit may only be operated with the filter being inserted.
Check the filter within regular intervals and exchange it if contaminated.
8.2
Errors and how to eliminate those
DISCONNECT FROM MAINS
Risk of fatal injury!
Unplug the unit before opening it.
ELECTRIC SHOCK
Danger of electric shock when unit is opened!
Error
Unit condition
Cause
Elimination
Unit cannot be
switched on
Unit has not been connected
Connect unit with mains supply
Defective fuse
Check fuse in the cold appliance plug and, if required,
exchange the fuse; if error occurs repeatedly, put unit
out of operation and inform service agent. See also
“
Incorrect mains voltage
”
Defective power cord
Replace power cord
Incorrect mains voltage
Use voltage selector to adjust correct input voltage
Initialization / Self
test
No chip card
Insert chip card
Chip card not readable
Chip card defect or chip card may have been
inserted incorrectly
Wrong chip card
Insert correct chip card
Polymerisation
process is not
started or aborted
Limit switch was activated
Close maintenance flap and / or drawer correctly.
If necessary, repeat polymerisation process
No flash lamp inserted or flash lamp
defect
Insert flash lamp and, if required, start polymeri-
sation process again; if error occurs repeatedly,
exchange flash lamp
Symbol explanation
No light appears
Permanent light
1 LED flash per second
2 LED flashes per second
Alternating flashing LEDs
7.6.3 Filter drawer
Has the filter been inserted?
Clean or replace very dirty filters.
7.6.4 Inside the machine
Has the flash lamp been inserted as far as it will go?
Has the chip card been correctly inserted into the card reader?
Is the filter disc in position and clean?
8 Appliance
8.1
Working with the machine
Connect the machine to the power supply and switch it on.
Make sure that the material is placed in the machine on the supplied object
holder (if possible!) and not with the whole plaster model. The large light
reflector allows this to be done, but it significantly prevents the cooling
capability.
Warming can be reduced with the “
Model tray
” spacer when the 180 second
light polymerisation is used frequently. For this purpose, the model must
be placed on the tray with the feet pointing downwards in order to achieve
better ventilation.
Delivery form:
2 pieces “
Model tray”
REF 66017525
Open the polymerisation drawer and insert the prosthetic material in the light
reflector. Close the drawer. Press the button for the respective polymerisation
cycle. The respective LED ring is lit.
The drawer is locked (not in the 6 second cycle). At the end of the polyme-
risation time an acoustic signal (buzzer) is heard and the drawer is unlocked
again. Remove the prosthetic restoration and exchange the reflector pot if it
is heavily contaminated or clean it.
The fan which started at the beginning of polymerisation continues to run until the interior of the unit has been cooled down
to a large extent.
NOTE
The built-in ventilator must cool the prosthetic material evenly and well to ensure an excellent polymerisation result.
This will only happen if the product is positioned in the centre.
Observe the following information:
Place the prosthetic material in the centre of the light reflector.
Align the polymerisation product with the light source to ensure an excellent polymerisation result. If polymerisation is
advantageous on the model, place the model with the greatest possible angle to the light reflector wall. This keeps the
ventilation holes in the floor free. An angular position is favourable as lateral exposure is better on the facing.
When positioning the product, ensure that it is aligned beneath the edge of the light reflector. Otherwise,
the drawer will not close and the object will touch the window pane in the machine.
Insert the drawer as far as it will go.
Select the required polymerisation time and press the button.
NOTE
The 6 second polymerisation time is used to harden the material. All components of Signum
®
(exception: Opaque)
can be pre-polymerised using the 6 second polymerisation time if a maximum layer thickness of 2 mm is not exceeded.
Then carry out intermediate polymerisation for 90 seconds and end polymerisation for 180 seconds.
To ensure extended service life of the unit, the interior of the unit should have been cooled down sufficiently before the
unit is switched off. The unit should not be switched off as long as the fan is not in standby-mode. (Fan is not turned
off completely).
Attempting to pull out the drawer may result in blocking of the locking magnet. In such cases the drawer needs to be
re-inserted and the unit must be switched off and on; then the drawer can be opened again.
Model tray
Summary of Contents for HiLite power
Page 1: ...DE Betriebsanleitung GB Operating instructions HiLite power DE GB...
Page 2: ......
Page 18: ...16 DE 14 Schaltplan 14 1 Klemmenbelegungsplan...
Page 20: ......
Page 36: ...16 GB 14 Circuit diagram 14 1 Terminal allocation scheme...
Page 38: ......
Page 39: ......