CONTRAINDICATIONS 

No known contraindications.

PRECAUTIONS 

Risk of allergic reactions can not be excluded for those predisposed to allergies. 

STERILIZATION, DISPOSAL & STORAGE

Leaves: One Time Use, Disposable 

The disposable Leaves are meant for 

single use only

. Due to the 

nature of the material and design of the Leaves, they should not 
be sterilized and reused. Leaf mouthpieces must be properly 
discarded in a clinical waste container to prevent transmission 
of infectious pathogens.

U-Shaped Connectors 

100% Autoclavable and Reusable

 

Prior to placement in an ultrasonic cleaner, rinse the U-Shaped 

Connectors with water and brush both external and lumen 
surfaces with a soft bristle brush for the exterior surface and a 

soft bristle lumen brush or pipecleaner for the interior surfaces. 

Place in an ultrasonic unit filled with enzymatic cleaner such as 

VibraKleen E

2

 (note: follow instructions with enzymatic cleaner for 

proper mixing, exposure time, etc. per manufacturers instructions). 

Rinse with water and place in an autoclavable pouch and put through 
a wrapped cycle in the autoclave (See Sterilization Cycle Table).

Short Hose 

(8" Quick-Disconnect Hose):  

100% Autoclavable and Reusable

 

Prior to placement in an ultrasonic cleaner, rinse the Quick Disconnect Hose 

with water and brush both external and lumen surfaces with a soft bristle brush for the 
exterior surface and a soft bristle lumen brush or pipecleaner for the interior surfaces. 
Place in an ultrasonic unit filled with enzymatic cleaner such as VibraKleen E

2

 (note: follow 

instructions with enzymatic cleaner for proper mixing, exposure time, etc. per manufactur

-

ers instructions). Rinse with water and place in an autoclavable pouch and put through a 
wrapped cycle in the autoclave (See Sterilization Cycle Table).

Long Hose 

(48" HVE Attachment Hose with Quick-Disconnect Adapter):  

100% Autoclavable and Reusable

 

The Long Hose should stay connected to the HVE until ready for routine evacuation  
cleaning with an enzymatic evacuation cleaner such as Kulzer VibraKleen E

2

Running evacuation cleaner through the hose to clean out any debris at least once  
per week is recommended. It is not necessary to run evacuation cleaner through  
the hose between each patient.

Storage 

Please ensure ReLeaf is stored at room temperature.

The Short Hose (8" Quick Disconnect Hose) is specifically 

designed to be placed in an available rack slot when not in 
use. (See right)

VALIDATED STERILIZATION CYCLES

Gravity Displacement Autoclave

Pre-vacuum Autoclave

TEMPERATURE: 121

°

C/250

°

TEMPERATURE: 132

°

C/270

°

F

MINIMUM EXPOSURE TIME: 30 minutes 

MINIMUM EXPOSURE TIME: 4 minutes

MINIMUM DRY TIME: 30 min

MINIMUM DRY TIME: 20 minutes