Setting up, installation and putting into operation
The instrument is to be mounted on a solid, non combustible surface (e.g. laboratory bench (H x W x D): 450 x 315 x 550 mm)
in such a manner as to achieve a secure horizontal position.
Ambient conditions:
Temperature range 0°C (32°F) to 40°C (104°F)
Relative humidity between 30% and 75%
Mains voltage ± 10% from the nominal value
Use only indoors
No direct exposure to UV light
Max. 2.000 m above sea level
Overvoltage category II
Pollution degree 2
Compressed air connection:
Connect to compressed air supply using the connection at the rear of the unit (
7
). Maximum pressure: 10 bar.
Note!
The compressed air must be clean and dry!
Any guarantee claims shall be excluded in case of malfunctions or damage resulting from
inadequate compressed air supply!
(See special accessories chapter 4.)
Mains connection:
Prior to operating, check whether the nominal voltage corresponds to the value indicated on the type plate
the unit. The unit must be connected to a contact socket with a ground. Only connect the device to a mains
socket with a ground connection.
Unit may only be operated with the enclosed power cord!
Be sure to comply with technical specifications when the mains cable is replaced. The sheathing of the mains
connection cable must be approved for a temperature of 90 °C (194 °F) or higher. See section 9. When using
country-specific mains cables, make sure that the cables are certified in accordance with the electrical rating
data and notes of the relevant standards of the respective country.