Working with the Palamat Premium
Attention!
Check the content of the shipping carton and compare with the scope of delivery described in the operating
instructions (see chapter 4 scope of delivery).
Prior to operation it must be verified whether the information provided on the type plate correspond with the local
power supply data.
The Palamat Premium polymerization unit features a cold-appliance plug and a safety plug. Power cord and plug
must be checked for damage prior to the use. If any damage is found, the unit must not be connected to the mains
supply.
The Palamat Premium polymerization must be connected to the mains supply via properly installed sockets with
protective conductor connections (protection class I). The user must have easy access to the plug and mains socket.
7.1
Pressure pot
To open the lid (
1
), locking lever must be in upright position. Turn the lid by 90°, tilt it and remove it from the pressure pot.
After completing the polymerization process, the handle must be in a perpendicular position to open the lid.
The lid can only be removed if it has been lowered into the pot after release of pressure.
The lid is inserted and closed in reverse order. It must ensured that the supporting strut of the lid fits into the grooves on the
pressure pot rim.
Opening and closing the pressure pot
1
3
2
1)
Place handle in perpendicular position and turn lid by 90°
2)
Tilt lid slightly
3)
Remove lid
The pot is closed in reverse order.
7.2
Filling with water
The unit must not be heated without water. Minimum filling level: 4 cm. Even if objects are placed in,
the filling level is always below the upper end of the air supply fitting.
Do not fill the unit under running water. Kulzer recommends filling the device using a suitable container
with a spout.
Prior to filling with water, check if the water discharge valve at the rear of the unit is closed.
Use only pure tap water for processing.
Distilled, deionized water or other liquids may not be used!