- 8 -
GB
DE
GB
FR
ES
IT
PT
NL
SE
DK
NO
FI
GR
7.2.1 Overtemperature protection
The Palamat Premium is equipped with two upper limit temperature protection fuses that are triggered when
the temperature exceeds 110 °C (230 °F).
In the event of activation of the fuses, the heating circuit is interrupted.
Check the minimum water level regularly to avoid overheating.
Note!
The overtemperature protections may only be reset by trained service personnel.
Contact your responsible service agent (see chapter 11 Service).
7.3 Polymerization
To switch on the unit, press the power switch (
5
) at the front of the unit. The power switch is lit and indicates that the unit
has been started up. The Kulzer welcome screen with the software version is briefly shown on the TFT display (3) start screen.
7.3.1 Edit program (customize individually)
The device has three freely configurable programs (A / B / C) that are green / blue / purple color coded. In order to enter the
editing area, one of the interactive buttons (15 / 16 / 17) on the Home display must be pressed for approx. 1.5 seconds. An
audible beep will sound. The screen then changes to the Edit display for the respective program.
Home display
In this area, the temperature (D), pressure (E) on / off (BAR) and time (F) can be configured individually. Press the interactive
button (18) to save the configuration settings. You can leave the editing area at any time by pressing the Home button (11).
In this case, the changed settings are not saved. If the Home button is pressed for one second, the automatic heating
function and standby temperature are deactivated.
Editing display
Temperature
The program temperature selection is confirmed by pres-
sing the button (15) on the Edit display. The display screen
changes to the Temperature setting.
The temperature is configured using the assigned plus &
minus buttons. The adjustable temperature range is 40°C –
100°C. Confirm the selection with the Home button (11).
Time
Selecting the program time is confirmed by pressing the
button (17) on the Edit display. The display screen changes
to the Time setting.
The time is configured using the assigned plus & minus
button. The adjustable time range is 0:01 – 9:59 [h:mm].
Continuous operation can also be set using the ’∞’ symbol.
Confirm the selection with the Home button (11).
D
11
E
18
B
A
C
15
16
17
18
F