9
Technical data
Power consumption
650 W
Rated voltage / fuse type
100 V, 2 x T8A / 250 V
120 V, 2 x T6.3A / 250 V
230 / 240 V, 2 x T4A / 250 V
Nominal frequency
50 / 60 Hz
Voltage tolerance for all mains voltages
+-10%
Mains cable
EU - H05V2V2-F 3G 0.75 250VAC/16A
US - SJT3X18AWGWB105C C13/2,5m black 125VAC/15A
Protective class
1
Pollution degree
2
Pressure hose connection
3 to 10 bars
Nominal pressure
2 bars
Max. operating pressure
2.8 bars
Operating temperature of water
40 – 100°C (113 – 212°F)
Temperature limit
Dev / - 2.5 °C (4.5 °F) or approx. 5%
Height
Width
Depth
Weight
Dimensions
Required space
Height 300 mm
approx. 450 mm
Width 305 mm
approx. 315 mm
Depth 400 mm
approx. 550 mm
Weight 8.6 kg
10
Information on malfunctions
Error messages
Error
Cause
Remedy
E 98
Sensor break / short circuit (heating plate) or
automatic software - Safety shutdown due to
high temperature
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
E 99
Sensor break / short circuit (water bath) or
automatic software safety shutdown due to
high temperature
Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
E 18
Heating defect: If there is no temperature
increase after a defined period of time.
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
E 19
Heating control: If the control temperature
exceeds the target temperature for a defined
period of time
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
11 Service
We appreciate your comments, feedback, and suggestions.
Contact in the countries
and more information are
available by the displayed QR code
or at our website
www.kulzer.com
Micro USB port: (See section 5)
The USB port is used for software updates, which may only be performed by trained service personnel.
Therefore, customers are not permitted to connect USB cables/adapters, etc.