- 11 -
GB
Function /
Error Condition
Hand piece Button operation
Acoustic signal/s emitted
Display
indication
BUILD-IN Tempera-
ture overload sensor
defective
•
4 beeps
at the start of
an exposure cycle.
blinking
Charging circuit of the
hand piece defective
Note:
This condition is
detected only with hand
piece placed inside the
charging unit.
•
1 beep
per second for the length
of 30 seconds.
blinking
“CLEANING MODE”
• With hand piece in the “
Sleep Mode
”
condition,
PRESS
and
KEEP PRESSED
SIMULTANEOUSLY
the “
TIME / PRO-
GRAM
” and the “
STAND-BY
” buttons
for 5 seconds.
•
1 beep
when the “
TIME / PRO-
GRAM
” and the “
STAND-BY
”
buttons are simultaneously
pressed.
•
1 beep
at the beginning of the
“
Cleaning time
”.
•
1 beep
at the end of the
“
Cleaning time
”.
Display is
switched
off .
4.2.6
LED information concerning the radiation emitted
WARNING
Photobiological safety of lamps and lamp systems IEC 62471. According to the Standard IEC 62471, the device
results in risk class 2 (moderate risk) concerning a retinal risk from blue light or thermal retinal risk. The above
warning indications are applied to the device package.
Risk Group 2
CAUTION. Possibly hazardous optical
radiation emitted from this product.
Do not stare at operating lamp. May
be harmful to the eye
Product tested against IEC62471
1707_11667_GBA_Translux_2Wave_EU_NENA.indb GB11
1707_11667_GBA_Translux_2Wave_EU_NENA.indb GB11
07.09.17 16:22
07.09.17 16:22