- 15 -
GB
Ending the pre-qualifi cation phase, without detecting any battery defect,
the “
regular
” battery charging process starts and all the four black bars inside the
battery icon blink progressively
(Fig. b).
IMPORTANT NOTE – Defective battery
If during the pre-qualifi cation phase, the hand piece electronics detects
a battery malfunction, the electronics interrupts the changing process and the
frame of the battery icon (without internal black bars) will blink (
Fig. c
).
In this mode, a
beep
signal is emitted every second for a period of 30 seconds.
Replace the defective battery with a new one!
WARNING
Use Kulzer batteries only!
b) Battery is partially charged
In case of a battery being partially charged, (one or more internal black bars are displayed inside the
battery icon) the charging process starts almost immediately after having inserted the hand piece into the
charging station.
All the four black bars inside the battery icon blink progressively (see Fig. b).
c) Battery is fully charged
When the charging phase is fi nished (battery fully charged) all the four black bars
inside the battery icon are permanently displayed (
Fig. d
).
NOTE
When the hand piece is placed in the charging unit its functioning is inactive.
The operativeness of the hand piece buttons are prevented.
The hand piece can be removed from the charging unit during the charging process; however, the battery
may not be fully charged and may shorten the operational duration.
6 Operation
WARNING – Check the condition of the device before starting the treatment.
Before each treatment, always make sure that the device is in proper working order and that the accessories
(light guide and protective light cone) are effi
cient. If anything is noted during treatment, do not carry out the
treatment and contact Kulzer or a local authorised Customer Service agent (see paragraph 14 Service).
Do not use the device and its accessories if damage is apparent. Prior to each use of the Translux 2Wave
ensure that the emitted light intensity is suffi
cient to guarantee the polymerisation. Check the light intensity by
using the light meter integrated into the charging unit (see paragraph 6.3 Measurement of the Light Intensity).
WARNING – Control of infections
First use
The accessories, light guide and protective light cone, are delivered in
NON STERILE
condition and must be
sterilised before being used for the fi rst time.
Every
use
Once used, the accessories, light guide and protective light cone, must be reprocessed prior to reuse,
according to the procedures described in the paragraph 8 (Cleaning, disinfection and sterilisation).
b
c
blinking
d