- 16 -
GB
6.1
Selecting the curing time
The Translux 2Wave can operate with four diff erent exposure cycles:
•
Exposure with slowly increasing output:
“
Soft-Start
”
(cycle duration 20 seconds).
The “
Soft-Start
” means increase of light intensity from 50% up to 100% within 2 seconds.
•
Exposure with constant output:
cycle duration 20 seconds, 10 seconds or 5 seconds.
Press
the
“
Time / Program
” button
2 seconds to activate selection and scroll through
the curing modes on the display.
The selected exposure time is indicated (in the display) by a round spot next to the selection.
NOTE
When the “
Soft-Start
” program is selected, round spots
simultaneously lighten next to each icon
and
:
Each time the “
Time / Program
” button is briefl y pressed, the display setting advances
to the next lower exposure curing mode in a loop.
NOTE
During the light emission the “
Time / Program
” button is inactive.
NOTE – “Freeze Mode”
10 seconds after selecting an exposure mode the “
Time / Program
” button is deactivated.
Press the “
Time / Program
” button 2 seconds to activate selection and scroll through the curing modes
on the display.
6.2
Activating and deactivating the exposure cycle
Briefl y press the “
STAND-BY
” button
. The light emission will turn on.
Once the curing cycle is complete, the unit will automatically turn the light off .
Acoustical signal:
The following acoustical signals (beep) are emitted:
1 beep
while press the “
STAND-BY
” button.
1 beep
after 10 seconds of exposure time (only 20 second Mode).
1 beep
at the end of the exposure time.
NOTE
The exposure cycle can be interrupted at any time, before the exposure time is over,
by pressing again the “
STAND-BY
” button.
Acoustical signal:
An acoustical signal (
beep
) is emitted, when the “
STAND-BY
” button is pressed to interrupt the cycle.
6.3
Measurement of the Light Intensity
The charging unit of the Translux 2Wave is equipped with an integrated light intensity meter (
8
).
CAUTION
Before each treatment always check the light intensity by using the light intensity meter
integrated in the Translux 2Wave charging unit.
• Without applying pressure, place the light guide fl at and directly on the testing area (
8
).
Activate the unit by pressing the “
STAND-BY
”
button (whichever exposure time selected).
The light intensity measuring is immediately indicated by the 2-colour LED (Symbol
) placed at the bottom right
of the testing area.
GREEN LED
: The light intensity is suffi
cient for the treatment.
YELLOW LED
: The light intensity is insuffi
cient and patient/s must not be treated.
CAUTION
If the light intensity is insuffi
cient, please carry out the following controls and measures:
– The light guide has not been inserted into the hand piece (see paragraph 5.4 Inserting the fi bre optic
into the hand piece).
– Check the light guide for contamination or defects.
– Clean the light guide of any contamination (see paragraph 8 Cleaning, disinfection and sterilisation)
or replace the damaged light guide with a new one.
Use original Kulzer light guides only.
2x
Soft-
Start
2x