6 | Function description
The [Management] tabulator with three dialog windows
6.12.3.1 The [Start teaching] button
To redefine the heating head positions, always perform function [Start device
setup]:
– whenever reinstalling the Rework System
– after reinstalling HRSoft 2
– after replacing the heating head
– after relocating the Rework System.
To do so, you will need the calibration kit item no. 324627, consisting of a square
black plate and a round focusing small plate. Process [Start device setup] is run be-
ing guided by an assistant, similar to the wizard guiding you through the soldering
process.
Use one of size 5 or 10 nozzles included in the scope of delivery for this process.
With regard to this, please also read Chapter Replacing the soldering nozzle
[
Info: To fully set up the Rework system after changing the heating head or relocat-
ing the system, also run the function associated with the [Adjust pixel shift] button
in the [Device options] dialog. With regard to this, please read Chapter The [Pixel
shift adjustment] button.
CAUTION
Risk of burns from hot surfaces!
a) Touching the device surfaces may cause burns!
a) Push the lever to switch between normal view and the telephoto lens of the
Visionbox through to the teleposition. With regard to this, please also read
Chapter Overview of the machine parts, no. 11.
b) Click on button [Start device setup].
ð
The device first carries out a reference run of the pipette and the heating head.
ð
The assistant for setting the soldering head height is started. The info line
above the diagram describes the work steps to be performed.
ð
You will be asked to hold the square black plate included in the calibration kit
under the heating head.
Ersa GmbH
Operating instructions HR 550 XL EN 3BA00247-01 | Rev. 1